A weighting rolls off low and very high frequencies to imitate the ear’s sensitivity at moderate levels, and it is the default for most environmental and workplace noise assessments. C weighting keeps more low frequency energy, so it is used for peak and high level checks and for judging low frequency rumble. Z weighting applies no correction at all and reports the raw physical spectrum, which is what you want when you need to know what the source actually emits rather than what it sounds like. Many inexpensive meters offer only A and C, and some offer no weighting switch at all; if your task requires Z weighting, check the listing before buying.
What a weighting curve actually does to the numbers
A sound level meter does not report a single number for a sound. It measures pressure fluctuations across a range of frequencies, then applies a correction curve before adding the energy together. That curve is the weighting. The three you will meet in consumer and trade instruments are A, C and Z.
A weighting attenuates low frequencies heavily and high frequencies slightly, following a curve that approximates the ear’s sensitivity at moderate loudness. At 1 kHz the correction is zero by definition. Below a few hundred hertz the correction becomes large, so a deep hum that is physically strong can register much lower on A weighting than on C. Above a few kilohertz A weighting also rolls off, though less steeply than at the bottom.
C weighting is much flatter. It still rolls off at the extremes, but in the low frequency region it holds closer to the true level. That makes it useful where the energy is concentrated low, or where you are checking whether a sound is loud enough to damage hearing rather than how annoying it sounds.
Z weighting, sometimes called zero or flat weighting, applies no correction within the meter’s stated band. The reading reflects the physical level at the microphone across the band. It is the least processed number the instrument can give you, and it is the one to quote when you are documenting a source rather than a subjective impression.
The practical consequence is that you cannot compare an A weighted figure with a C weighted figure and treat the difference as noise. They are answers to different questions. A meter that displays both is telling you two different things about one event.
Why the same noise gives three different readings
Imagine a ventilation plant running in a small plant room. The fan produces a broad hum with most of its energy below 250 Hz, plus some blade tone higher up. On A weighting the low hum is discounted, so the meter might read in the low seventies. Switch to C weighting and the same event might read several decibels higher, because the low frequency energy now counts. Switch to Z weighting and the figure can climb again, depending on how much energy sits outside the A curve’s pass band.
The size of the gap is not fixed. It depends entirely on the spectrum. A hiss from a leaking compressed air line has most of its energy high, so A and C readings will sit close together. A diesel generator or a large transformer has most of its energy low, so the gap widens. This is why an A weighted reading alone tells you almost nothing about whether low frequency energy is present.
That matters in real work. A worker complaining about a headache and fatigue in a plant room may be exposed to low frequency energy that an A weighted survey underplays. A neighbour complaining about a distant industrial hum may be reacting to a component that A weighting barely registers. In both cases the A figure is not wrong; it is simply answering the question of how loud the sound seems, not how much low frequency energy is arriving.
Matching the weighting to the job in front of you
Most environmental and occupational noise limits are written in A weighted decibels, often shown as dB(A). If you are checking against a limit, the limit document tells you which weighting to use, and you should use that one regardless of what the meter defaults to. A reading in the wrong weighting is not a near miss; it is a different quantity.
C weighting comes into its own for peak measurements and for high level checks. Many hearing protection and impulse noise assessments use C weighted peak, because a short, sharp event can carry energy that A weighting smooths over. If you are measuring a press, a nail gun or a door slam, the peak figure is usually a C weighted or Z weighted peak, not an A weighted average.
Z weighting is the choice when you need to characterise a source or a space rather than judge it. Acoustic consultants use it when feeding data into a model, when checking a meter against a calibrator, or when documenting a low frequency problem. It is also the setting to use if you intend to apply your own correction later, because it preserves the raw level.
One practical note: many low cost meters offer A and C only. The weighting switch is often a small button marked A/C, and the display shows which is active. If the listing does not mention Z weighting, assume it is absent. The Extech 407730 sound level meter is an example of a meter that publishes A and C weighting and a 40 to 130 dB range, with 2 dB accuracy and 0.1 dB resolution; its listing does not mention Z weighting, so treat it as an A and C instrument.
The mistake that makes a weighting comparison meaningless
The commonest error is to take one reading on A, one on C, and present the difference as if it were a property of the source. It is not. The difference between A and C weighted readings is a function of the spectrum and the level, and it changes as the source changes. A single pair of numbers taken minutes apart, or on different meters, cannot be subtracted to give a meaningful low frequency correction.
The second error is to assume that a meter with an A/C switch is measuring both at once. Most inexpensive meters measure one weighting at a time, selected by the switch. If you need both figures for the same event, you need a meter that stores or displays both, or you need to repeat the measurement under identical conditions. The listing should say which. Where it does not, assume one at a time.
The third error is to treat the weighting as a minor display preference. It is not a preference; it is part of the measurement definition. A report that says 78 dB without stating the weighting is incomplete, because the same event could be 78 dB(A) and 86 dB(C). Always record the weighting alongside the number, and state the time weighting (fast, slow or impulse) as well, since that changes the figure too.
Where weighting choices stop mattering because the instrument is wrong
Weighting only applies to instruments that measure sound pressure across a frequency range. It does not apply to air quality monitors, gas detectors or particle counters, even when those instruments use the word weighting in their marketing. A CO2 monitor reporting parts per million is not applying an A, C or Z curve; it is reporting a concentration. If a listing for an air quality product mentions weighting, check what it is actually weighting, because the term is sometimes used loosely.
The Dienmern 8-in-1 air quality monitor and the Dienmern 5-in-1 CO2 monitor both report CO2, temperature and humidity, and the 5-in-1 listing states a 400 to 5000 ppm range with plus or minus 50 ppm accuracy. Neither listing mentions sound measurement or frequency weighting, so neither is a substitute for a sound level meter. The Aolidsive wall mount air quality monitor uses an NDIR CO2 sensor and reports temperature and humidity; again, no sound weighting applies. These are concentration instruments, and the accuracy question for them is covered in CO2 monitor accuracy explained.
Similarly, a sound level meter is not an alarm and not a hearing protector. It tells you a level; it does not stop the exposure. If your task is to protect people, the meter is one input to a decision, not the decision itself.
Reading a specification sheet without being misled
When a listing states a range such as 40 to 130 dB, that range is usually quoted for a particular weighting, most often A. The same meter may have a narrower range on C or Z, or may not offer those weightings at all. The listing should say. If it does not, ask before you rely on the top or bottom of the range.
Accuracy figures need the same care. A stated 2 dB accuracy is a tolerance, not a guarantee that every reading is within 2 dB of truth; it usually applies under reference conditions and across part of the range. A 0.1 dB resolution is a display convenience, not an accuracy claim. A meter that shows one decimal place is not thereby accurate to one decimal place.
The table below summarises what each weighting is for and what to check on a listing.
| Weighting | What it approximates | Typical use | What to check on a listing |
|---|---|---|---|
| A | Ear sensitivity at moderate levels | Environmental and workplace noise limits | Stated range and accuracy at A weighting |
| C | Flatter response, more low frequency retained | Peak and high level checks, low frequency assessment | Whether C is selectable and what the range is on C |
| Z | No correction within the meter band | Source characterisation, calibration checks, modelling input | Whether Z is offered at all; many budget meters omit it |
If the listing is silent on Z weighting, write “Not stated in the listing bullets” in your notes and proceed as if it is unavailable. That is the honest position, and it prevents you from planning a measurement the instrument cannot make.
The short version for a buyer
If your task is a noise limit or a general environmental survey, you need A weighting and you should confirm the meter offers it with a stated range. If you are checking peaks or low frequency, you need C weighting as well, and you should confirm the range on C. If you are characterising a source or feeding data into a model, you need Z weighting, and you should not assume a budget meter has it. The weighting is part of the measurement, so record it with every number you write down. And if the instrument in front of you measures CO2 or air quality rather than sound pressure, weighting does not apply at all; that instrument answers a different question.
Frequently Asked Questions
Can I convert an A weighted reading to a C weighted reading?
No, not from a single number. The difference between A and C weighted levels depends on the frequency spectrum of the sound, and that spectrum is not recoverable from one A weighted figure. You need either a meter that reports both, or a spectrum measurement.
Is Z weighting the same as unweighted?
In practice, yes, within the meter's stated frequency band. Z weighting applies no correction across that band, so the reading reflects the physical level at the microphone. Outside the band, the meter's response still rolls off, so Z is not infinite bandwidth.
Why does my meter show a different number when I press the A/C button?
Because the two weightings apply different correction curves to the same sound. The gap is largest when the sound has strong low frequency content. It is not a fault; it is the meter doing what the weighting is designed to do.
Do I need Z weighting for a workplace noise assessment?
Usually not for the compliance figure, which is normally A weighted. Z weighting becomes useful if you suspect a low frequency problem or need to characterise the source. Check the assessment method you are following before deciding.
Does a CO2 monitor use A, C or Z weighting?
None of them. A CO2 monitor reports a gas concentration in parts per million, not a sound pressure level. Weighting curves apply to sound measurement. If a listing uses the word weighting for an air quality product, check what it is actually describing.
What if the listing does not say which weighting the accuracy figure applies to?
Then the figure is incomplete for your purposes. Write "Not stated in the listing bullets" and treat the accuracy claim as unverified for the weighting you intend to use. A tolerance quoted without a weighting and a level is not a specification you can plan around.