How to read an accuracy figure written as percent plus digits

A figure such as plus or minus two percent plus two digits looks like one number and behaves like two. The percentage follows the reading or the span, while the digits term stays the same size no matter how small the measurement gets. This article shows how to split the figure, work it out at the reading you care about, and what to do when a listing publishes only half of it.

As an Amazon Associate, Akermin earns from qualifying purchases. Product links on this page are affiliate links.

An accuracy figure written as a percentage plus a number of digits is two separate error terms added together, and you have to work out both at the reading you actually care about. The percentage is proportional, so it grows with the measurement or with the span it is tied to, while the digits term is a fixed number of counts of the last displayed place and never changes size. At the top of a scale the percentage dominates and the digits are noise. At the bottom the digits dominate completely, which is why a specification that looks tight at full scale can be meaningless on a small reading.

The two halves of a plus or minus figure

Write the figure out as a sum and it stops being mysterious. Plus or minus two percent plus two digits means take two percent of something, take two counts of the smallest place on the display, and add them. The result is a band around the displayed value. The true value is claimed to sit somewhere inside that band, not at its centre and not at any particular point in it.

The two terms exist because instruments make two different kinds of error. Gain errors scale with the quantity being measured: a reference voltage that sits a fraction high makes every reading a fraction high, and the absolute size of that mistake grows as the reading grows. Offset errors do not scale at all. A converter whose zero sits a count or two away from true adds the same small amount whether the sensor sees almost nothing or almost everything. One percentage cannot describe both honestly, so the specification carries one term for each.

This is also why the same instrument can be described as excellent and as useless depending on which end of its range you quote. The band in absolute units is widest at the top. The band as a fraction of what you are reading is widest at the bottom, and that fraction is the number that decides whether a measurement tells you anything.

What a digit actually is on a given display

A digit, sometimes written as a count or as LSD, is one step of the least significant place the display can show. On a readout that resolves to one ppm, two digits is two ppm. On a readout that resolves to hundredths of a pH unit, two digits is two hundredths. The phrase is identical in both specifications and the quantity it stands for differs by a factor of a hundred, which is why a digits term can never be read without first finding the resolution it attaches to.

That link between the two numbers trips people up in both directions. A manufacturer who improves the display from one place to two places and leaves the digits term unchanged has quietly tightened the specification by a factor of ten. A manufacturer who widens the span without adding a display place has loosened it by the same mechanism without changing a word of the text.

Product names make the confusion worse by printing a resolution where a reader expects an accuracy. The VIVOSUN digital pH and TDS meter carries the phrase describing hundredths in its product name while the bullet that discusses the three in one TDS function states an accuracy of plus or minus two percent across a measurement range of zero to nine thousand nine hundred and ninety ppm and zero to nine thousand nine hundred and ninety microsiemens per centimetre. A step size of a hundredth and a band of two percent are answers to two different questions, and only one of them is an accuracy claim.

Percent of reading and percent of full scale are not the same promise

A percentage has to be a percentage of something, and in instrument specifications it is one of two things. Percent of reading means the proportional term shrinks as the measurement shrinks, so the band follows the number on the display down. Percent of full scale means the proportional term is calculated once, from the top of the range, and then stays fixed for every reading the instrument will ever show.

On a wide span the difference is enormous. Two percent of reading on a measurement of twenty five ppm is half a ppm. Two percent of full scale on the same instrument, where full scale is nine thousand nine hundred and ninety ppm, is just under two hundred ppm. Same words, same number, two results that differ by a factor of four hundred on exactly the same measurement.

The honest conclusion is uncomfortable. When a bullet states a bare percentage with no scale attached, as the VIVOSUN bullet does, the reader cannot tell which promise is being made. The conservative reading is to assume full scale until a manual says otherwise, because that is the weaker of the two claims and the one a marketing bullet is more likely to be hiding.

A worked example across four points on one scale

Take a conductivity style instrument with a span of zero to nine thousand nine hundred and ninety ppm and a display that resolves to one ppm. Apply the same written specification twice, once interpreted as percent of reading plus two digits and once as percent of full scale, and watch what happens to the useful part of the measurement.

Reading Two percent of reading Two digits Total band as percent of reading Same words read as two percent of full scale
25 ppm 0.5 ppm 2 ppm 2.5 ppm, which is 10 percent 199.8 ppm, which is 799 percent
250 ppm 5 ppm 2 ppm 7 ppm, which is 2.8 percent 199.8 ppm, which is 80 percent
1000 ppm 20 ppm 2 ppm 22 ppm, which is 2.2 percent 199.8 ppm, which is 20 percent
5000 ppm 100 ppm 2 ppm 102 ppm, which is 2.04 percent 199.8 ppm, which is 4 percent

Two things fall out of the table. Under the percent of reading interpretation the digits term is irrelevant above a few hundred ppm and decisive below fifty, which tells you exactly where on the scale the instrument stops being a measuring device and starts being an indicator. Under the full scale interpretation there is no useful bottom to the range at all, and the instrument only earns its specification in the upper part of its span.

Accuracy figures published by four environmental monitors

Run the same question across a set of real listings and the pattern is consistent: ranges and sensor types are published, bands are usually not. The table below records what each set of bullets actually states, with no inference.

Instrument Accuracy band in the bullets Scale the percentage applies to What the bullets do publish
VIVOSUN digital pH, TDS and EC meter Plus or minus 2 percent on the TDS function Not stated in the listing bullets pH range 0 to 14, TDS range 0 to 9990 ppm and 0 to 9990 microsiemens per centimetre, temperature 0.1 to 80.0 degrees Celsius, protective case, not for sea water
BREATHE Airmonitor Plus Gen 2 Not stated in the listing bullets Not stated in the listing bullets Eight channels covering carbon dioxide, PM1, PM2.5, PM10, TVOC, formaldehyde, temperature and humidity, NDIR carbon dioxide sensing, laser particle sensing, ninety day history, six hour battery, data export
HENRUN portable particle counter Not stated in the listing bullets Not stated in the listing bullets Particles measured down to 0.3 microns, preset measurement modes, automatic data logging, rechargeable battery
Airthings View Plus Not stated in the listing bullets Not stated in the listing bullets Radon, PM2.5, carbon dioxide, VOCs, humidity, temperature and pressure, electrochemical sensor type, operating temperature 4 to 40 degrees Celsius, app required

Marketing adjectives fill the gap where a band should be. The HENRUN particle counter describes laboratory grade precision and advanced sensor technology, which are not quantities and cannot be worked out at a reading. The Airthings View Plus publishes an operating temperature window of four to forty degrees Celsius, which is useful and is frequently mistaken for the narrower window inside which a band would hold. Reviews on this site treat the same gap the same way, whether the subject is the Temtop M10 Plus, the Temtop S1 Plus or the PurpleAir Classic.

Turning an accuracy figure into a sentence about your reading

Work in a fixed order and the figure stops being abstract. First find the resolution, because the digits term is meaningless without it. Second, find out whether the percentage is of reading or of span, and if the listing does not say, assume span and look for the manual. Third, pick the reading you actually expect to see in use rather than the top of the range, and compute both terms at that value. Fourth, add them and express the result as a fraction of your reading, which is the number that decides whether the measurement can support the decision you want to make with it.

Then ask what the band was quoted against. A reference temperature, a warm up period and a calibration interval are all part of the claim, and an instrument used outside any of them carries a wider band than the printed one. That discipline applies to a thermal instrument such as the TopTes PT520A exactly as it applies to a water quality pen, because the arithmetic belongs to the specification rather than to the sensing technology.

Frequently Asked Questions

Does a finer resolution mean a more accurate instrument?

No. Resolution is the smallest change the display can show, and accuracy is how far the displayed value may sit from the true value. A monitor that shows hundredths can still carry an error band many times wider than its last digit. The two numbers are set by different parts of the design, the display and converter on one side and the sensor and reference on the other, and a listing can be generous with one while saying nothing about the other.

Which is worse at a low reading, percent of reading or percent of full scale?

Percent of full scale, by a wide margin. A percentage of the span is a fixed quantity, so it stays the same size whether the display shows a tenth of the scale or all of it. Two percent of a nine thousand nine hundred and ninety ppm span is about two hundred ppm, which is a rounding detail at the top of the scale and larger than the entire measurement near the bottom.

What does a single accuracy figure mean on a multi function meter?

Usually that it belongs to one function and has been printed as though it covered them all. A combined pH, conductivity and temperature instrument has three separate sensing paths with three different error sources, and no single percentage can describe all three. When one figure appears beside a list of functions, treat it as applying to the function it is printed next to and nothing else until the manual separates them.

Can I treat a published band as a worst case?

Only within the conditions attached to it. A published band normally applies at a stated reference temperature, over a stated part of the range, after a stated warm up, and with the instrument inside its calibration interval. Outside those conditions a second term, often a temperature coefficient per degree, is added on top. The band is a worst case inside the box the manufacturer drew around it, not everywhere.

Why do some specifications add digits and others do not?

Because the digits term describes errors that do not scale with the measurement, such as converter offset, noise in the last place and the rounding of the final count. Instruments whose offset errors are genuinely tiny compared with the span can fold them into a single percentage without misleading anyone. When a specification carries no digits term and no separate offset term, the published band at the very bottom of the range is almost always optimistic.

Does the same band hold across the whole temperature range?

Rarely. Most specifications are quoted at a narrow reference temperature band and then add a coefficient for every degree outside it. An instrument rated for operation from four to forty degrees Celsius is telling you where it will work, not where it will hold its published accuracy. Those are two different ranges and the second is almost always narrower than the first.