Divide the milliohm value by 1,000, which is the same as moving the decimal point three places to the left. So 250 milliohms is 0.25 ohms, 47 milliohms is 0.047 ohms, and 3.2 milliohms is 0.0032 ohms. To go back, multiply ohms by 1,000. The most common mistake is confusing the lowercase m in mΩ, meaning one thousandth, with the capital M in MΩ, meaning one million.
The prefix behind the conversion
The ohm is the SI unit of resistance. Metric prefixes scale it in steps of a thousand, the same way millimetres relate to metres. Milli means one thousandth, so one milliohm (1 mΩ) is 0.001 Ω, and there are 1,000 mΩ in one ohm. Micro means one millionth: 1 µΩ is 0.000001 Ω, and 1,000 µΩ make 1 mΩ.
The formulas are:
- ohms = milliohms / 1,000
- milliohms = ohms x 1,000
- milliohms = microohms / 1,000
Upward from the ohm, kilo is a thousand and mega is a million. Insulation testers report megohms, where bigger is better; low resistance meters report milliohms, where smaller is usually better. The two sit about nine orders of magnitude apart, so mixing up m and M is not a small slip.
Conversion table and worked examples
| Milliohms (mΩ) | Ohms (Ω) | Where you might see it |
|---|---|---|
| 0.1 mΩ | 0.0001 Ω | Busbar joint or large crimp |
| 1 mΩ | 0.001 Ω | Current shunt resistor |
| 5 mΩ | 0.005 Ω | Internal resistance of a large battery cell |
| 25 mΩ | 0.025 Ω | Short run of heavy cable |
| 100 mΩ | 0.1 Ω | Typical multimeter test lead pair |
| 500 mΩ | 0.5 Ω | Motor winding or small coil |
| 1,000 mΩ | 1 Ω | Low value resistor |
| 2,200 mΩ | 2.2 Ω | Low value power resistor |
Example 1: a cell datasheet lists internal resistance as 18 mΩ. 18 / 1,000 = 0.018 Ω.
Example 2: a bonding test must be below 0.05 Ω and your meter shows 37.6 mΩ. 37.6 / 1,000 = 0.0376 Ω, which passes.
Example 3: a specification calls for 0.2 Ω maximum and you want it in milliohms for the meter’s range. 0.2 x 1,000 = 200 mΩ.
Counting zeros is where errors creep in. Write the decimal point explicitly: 0.0376, not .0376, and check your answer has the right number of places by converting back.
A spreadsheet helps when you have a column of readings. Put the milliohm values in one column and use a formula that divides each cell by 1000 in the next. Format the ohm column to four or five decimal places, because a default two place format will turn 0.0038 into 0.00 and hide the value completely. When you report results, keep the unit the reader expects; many test standards and maintenance records for joints and bonds are written in milliohms precisely because the ohm values have so many leading zeros.
Why milliohm values matter in practice
Small resistances become important when current is large. Ohm’s law gives the voltage drop, V = I x R, and the heat, P = I squared x R, and both must use ohms, so convert first.
Suppose a battery cable lug has a resistance of 0.8 mΩ and carries 150 A when a starter motor or inverter runs. Converting gives 0.0008 Ω. Voltage drop is 150 x 0.0008 = 0.12 V. Heat is 150 x 150 x 0.0008 = 18 W, concentrated in one small joint. A corroded lug at 4 mΩ would drop 0.6 V and generate 90 W, enough to discolour insulation. That is why electricians and battery technicians measure joints in milliohms instead of just checking continuity.
Other places milliohms show up: motor and transformer winding checks, circuit breaker contact resistance, PCB traces, welding cables, and the internal resistance of lithium cells. In each case, a change of a few milliohms can be the first sign of a problem.
Reading low resistance meters without mistakes
A general multimeter is the wrong tool for most milliohm work. Its test leads alone add something like a tenth of an ohm or more, and its resolution on the lowest resistance range is often 0.1 Ω, which is 100 mΩ, larger than many values you are trying to read. A dedicated four wire milliohm meter, such as the Extech 380560 milliohm meter or the AideTek VC480C milliohm meter, passes a test current through one pair of leads and measures voltage with a separate pair, so lead and contact resistance drop out. The approach is explained in milliohm meters for low resistance testing.
- Check which range the meter is on and read the unit symbol on the display. A reading of 125.4 on a 200 mΩ range is 125.4 mΩ, or 0.1254 Ω.
- Short the Kelvin clips together and use the zero or relative function if the meter has one.
- Clean the contact points. Oxide and paint add resistance that is real but not the resistance you want to measure.
- Place the voltage sense contacts inside the current contacts, as close to the joint as you can.
- Let the reading settle, record it with its unit, then convert.
If you only have a multimeter, short the leads, note the reading, and subtract it. That gets you a rough figure down to a few hundred milliohms, not lower. For grounding systems the method differs again; see earth ground tester versus multimeter readings.
Common errors and safety when measuring
| Mistake | Effect | How to avoid it |
|---|---|---|
| Reading mΩ as MΩ or the reverse | Error of a billion times | Read the symbol case, check whether the meter is a low resistance or insulation tester |
| Moving the decimal two places, not three | Value ten times too big or small | Convert back to check |
| Including lead resistance | Reading too high by the lead value | Four wire meter, or subtract shorted lead reading |
| Measuring a live circuit | Meaningless reading, damaged meter | Isolate and prove dead first |
Resistance meters inject their own small current and must only be used on circuits that are switched off and isolated. Large batteries are an exception people get wrong: you cannot measure a battery’s internal resistance with an ohms range, because the battery’s own voltage drives current into the meter. Use a battery internal resistance tester. On mains equipment, isolate, lock off and prove dead with a voltage tester before connecting any resistance meter, and leave insulation resistance testing, which applies high voltage, to someone trained in it; the difference is covered in insulation tester versus continuity test.
Quick check before you report a value
Write the meter reading with its unit, divide by 1,000 if it is in milliohms, then multiply your answer by 1,000 to make sure you land back on the original number; if a specification is in ohms and your meter reads milliohms, do the conversion on paper before deciding pass or fail.
Frequently Asked Questions
How many milliohms are in one ohm?
There are 1,000 milliohms in one ohm. So half an ohm is 500 milliohms and a tenth of an ohm is 100 milliohms.
Is mΩ the same as MΩ?
No. Lowercase m means milli, one thousandth, while capital M means mega, one million. One megohm is a billion milliohms, so confusing them gives a completely wrong result.
How do I convert microohms to ohms?
Divide microohms by 1,000,000. For example, 350 microohms is 0.00035 ohms, which is also 0.35 milliohms.
Why does my multimeter show 0.2 ohms with the leads shorted?
That is the resistance of the test leads, probe tips and internal connections. It is normal for a general multimeter. Subtract it from readings, or use a four wire milliohm meter for low values.
What is a good milliohm reading for a cable joint?
There is no universal figure because it depends on conductor size and joint type. Compare the joint with the same length of unbroken conductor or with similar joints in the same installation; a joint much higher than its neighbours needs attention.