What a thermal camera does when you cannot change its emissivity setting

A camera with no emissivity control has not removed the assumption, it has frozen it, almost always near the value that suits building materials. This article works out what that costs on a low emissivity surface, explains the second setting that even adjustable cameras rarely expose, and separates the jobs the instrument still does properly from the ones it cannot.

As an Amazon Associate, Akermin earns from qualifying purchases. Product links on this page are affiliate links.

Nothing is removed, the assumption is simply frozen. A camera with no emissivity control still has to convert collected radiation into a temperature, and to do that it applies a fixed value in firmware, almost always near 0.95 because that suits the building materials most owners point it at. On plaster, brick, timber, paint and cable insulation that assumption is close enough to be invisible. On bare metal it produces readings tens of degrees low, and because the setting cannot be changed the error cannot be corrected inside the instrument. What remains available is changing the target rather than the camera, comparing surfaces rather than reading them, and recomputing from saved radiometric data afterwards.

What the firmware assumes when you cannot change it

The conversion from detector signal to displayed temperature requires a number for emissivity, because the signal is a sum of emitted and reflected radiation and the split between them cannot be determined from the measurement alone. There is no mode in which that number is absent. A camera without a menu entry has the number compiled in.

Two values are common. The first is 0.95, chosen because it sits in the cluster that covers nearly every non metallic surface, and because it makes the instrument correct on the things a general buyer will point it at. The second is 1.00, which treats every surface as a perfect emitter and makes the displayed value the apparent blackbody temperature. The difference between the two assumptions is small on a high emissivity target and significant on anything else, so it is worth knowing which one you have.

Listings very rarely say. That means the honest position for a buyer is that the fixed value is unknown rather than that it is 0.95, and the place to resolve it is the manual. Where the manual is also silent, imaging a known high emissivity surface alongside a contact measurement of the same spot tells you empirically whether the firmware assumption sits near the top of the range. The adjustable case, and what to do with the control when it exists, is covered in setting emissivity on a thermal camera.

How far a locked value pushes a reading on bare metal

The size of the problem is calculable, and calculating it is more useful than a general warning. Take a surface genuinely at 60 degrees Celsius in surroundings at 22 degrees, which is a routine electrical inspection scenario, and a camera whose firmware is locked at 0.95. The table shows what it displays as the actual surface emissivity drops away from that assumption.

Actual surface emissivity Example surface Displayed temperature Shortfall against the true 60 degrees
0.95 Matt paint, plaster, tape, cable insulation 60.0 degrees Celsius None
0.60 Anodised or heavily oxidised metal About 47.5 degrees Celsius About 12.5 degrees
0.30 Rolled stainless steel About 35.5 degrees Celsius About 24.5 degrees
0.10 Bright aluminium About 26.6 degrees Celsius About 33.4 degrees
0.05 Polished copper About 24.3 degrees Celsius About 35.7 degrees

The arithmetic illustrates the mechanism rather than predicting any particular display, because a real detector responds over a band rather than the whole spectrum and none of these listings states its band. The pattern, though, is robust. By the time the real emissivity is down at 0.10, a surface hot enough to matter is displaying about four and a half degrees above the room, and the camera will not flag it under any threshold a user would reasonably set.

Notice also where the curve is steep. Most of the damage is done between 0.95 and 0.30; the last two rows add relatively little because the reading has already collapsed towards the ambient. This is why the distinction that matters in the field is not fine grained. It is whether the surface is a high emitter or not.

The second setting almost nobody exposes

Emissivity is only half of what the conversion needs. The other half is the reflected apparent temperature, the effective temperature of everything the target is bouncing towards the lens. Professional instruments expose it as a separate entry. Consumer cameras almost never do, including many that allow emissivity to be changed.

On a high emissivity surface this hardly matters, since the reflected term is a small share of the total and an error in it is diluted accordingly. On a low emissivity surface it becomes the dominant input. A correction applied with the right emissivity and a wrong reflected temperature can be as far out as no correction at all, which is why entering 0.05 into a camera and trusting the result is a mistake even when the control exists.

The practical implication for a locked camera is that it is missing the less important of the two assumptions. Changing emissivity without being able to set the reflected temperature improves a bad reading without making it trustworthy. That is one reason the target modification approach, giving the camera a high emissivity surface to look at, outperforms arguing with settings: it makes both assumptions nearly irrelevant at once.

Correcting the number after the fact

Some of this can be recovered later, and the distinction that decides whether it can is whether the saved file holds radiometric data or just an image.

A radiometric file stores a temperature or a radiance value for every pixel along with the parameters used to compute it. Analysis software can then substitute a different emissivity and a different reflected temperature and recompute the whole frame, which means a survey taken with the wrong assumption is salvageable provided the assumptions are known afterwards. Several listings in this category publish exactly this capability. The FOXWELL RT280 states transfer to a Windows PC with software supporting point, line and area temperature analysis, two and three dimensional views and automatic report generation, backed by 8 GB of storage for more than 20,000 images, and the DCTOPT camera states a detachable memory card holding over 10,000 images with built in analysis across points, lines and areas.

A flat image file cannot be corrected. The conversion has already been applied and the underlying values discarded, so what remains is a coloured picture. This is the single strongest argument for checking what format a camera saves in before buying it, and it matters more on a camera with no emissivity control than on one with it, because post processing is the only route left.

Two habits make later correction possible. Record the ambient temperature at the time of the survey, since the reflected temperature is usually close to it indoors. And record what the camera was assuming, even if the answer is that the assumption is unknown, because a note saying the firmware value was not published is still information the next person needs.

What these four listings publish about the control

The listings in this topic illustrate how thin the published information is, and how careful the conclusion has to be.

Listing Emissivity control Saved data and analysis
Thermal Master Thor001 Adjustable emissivity, stated in the bullets Nine point measurement, five imaging modes, nine palettes, detachable macro lens
FOXWELL RT280 Not stated in the listing bullets 8 GB storage, over 20,000 images, PC software with point, line and area analysis and report generation
DCTOPT Not stated in the listing bullets 4 GB detachable card, over 10,000 images, built in point, line and area analysis
MiLESEEY TR20 Not stated in the listing bullets Video recording, PC analyser software

Only the Thermal Master Thor001 names the control, in the bullet that also gives its stated plus or minus 2.7 degrees Fahrenheit accuracy and its wide measurement span. For the other three the correct statement is that the control is not stated in the listing bullets. That is not the same as saying it is absent. Manufacturers routinely leave settings out of a bullet list that exists to sell a product, and a menu entry is not a selling point in the way a pixel count is.

What follows for a buyer is procedural rather than evaluative. If the emissivity control matters for your work, treat its absence from the bullets as an open question and get it answered from the manual or the seller before purchase. What the control actually does, and the physics it addresses, is set out in emissivity in thermal imaging. Where a camera turns out not to have it, a single point instrument with a published spot size can sometimes be the better tool for the specific measurement, a trade covered in thermal camera against infrared thermometer.

Working inside the limit instead of around it

A camera with a frozen emissivity value is not a crippled instrument. It is an instrument with a narrower domain, and most of its domain is intact.

  • Fabric and insulation surveys are unaffected. Nearly every surface involved sits within a few percent of the firmware assumption, so the readings stand as they are.
  • Comparison still works everywhere. Identical surfaces carry identical errors, so the hottest of three matching conductors is the anomaly regardless of the absolute values. This is the technique that survives every emissivity problem.
  • Change the target rather than the setting. A matt tape patch or a flat painted area gives the camera a surface that matches what it already assumes, which fixes the reading without touching a menu.
  • Trend the same spot over time. A series of readings from one marked location under comparable load is meaningful even when the absolute number is wrong, because the error is constant.
  • Know when to stop. A bare metal surface, an absolute number needed for a report, and no way to apply a target is the point at which a contact measurement replaces the camera rather than supplements it.

The one thing not to do is to treat a displayed number from bright metal as a measurement because the camera offered no way to question it. Where the resolution of the image is the limiting factor instead, that is a separate and better understood problem, covered in thermal camera resolution explained.

Frequently Asked Questions

How do I find out what value my camera is locked to?

Look in the manual rather than the listing, and look for a specification line rather than a feature bullet. Where the documentation is silent, a practical test is to image a surface of known high emissivity and a contact measurement of the same spot together; if they agree closely, the fixed value is near the top of the range. That tells you what you need for interpretation without requiring the manufacturer to publish anything.

Does a locked camera still work for building surveys?

Yes, and this is the reason the limitation matters less than it sounds. Plaster, brick, concrete, timber, matt paint and insulation all sit close to the value a locked camera almost certainly assumes, so fabric survey readings are barely affected. The limitation bites on bare metal, which is mostly electrical and mechanical work.

Can software fix the reading afterwards?

Partly, and only if the software has the radiometric data rather than a picture. A saved image with per pixel temperature values can be recomputed for a different emissivity, which is why analysis software is worth having. A flat image file cannot be corrected at all, because the conversion has already been applied and discarded.

What is reflected apparent temperature?

It is the effective temperature of everything the target is reflecting towards the camera, and it is the second input the radiance arithmetic needs. On a high emissivity surface it hardly matters. On a low emissivity surface it dominates, which is why a correction that uses the right emissivity and the wrong reflected temperature can still be badly out.

Does an infrared spot thermometer have the same problem?

Usually a worse version of it. Many handheld spot thermometers fix emissivity at a single value and do not say so, and they also average over a spot whose size grows with distance, so a small target gets mixed with its surroundings. The failure mode is identical, and the absence of an image makes it harder to notice.