How Do I Find the Internal Resistance of a Multimeter in Both Voltmeter and Ammeter Modes?

You want to know how much your multimeter loads a circuit, either as a voltmeter across it or as an ammeter in series with it. Both numbers can be measured at home with a battery, a resistor and a second meter.

As an Amazon Associate, Akermin earns from qualifying purchases. Product links on this page are affiliate links.

For voltmeter mode, put a known high-value resistor in series with the meter across a steady DC source, compare the reading with the source voltage, and calculate input resistance as R times Vm divided by (Vsource minus Vm); most digital meters come out near 10 megohms. For ammeter mode, pass a known current through the meter and measure the voltage across its input jacks with a second meter; the ammeter resistance is that burden voltage divided by the current, and it changes with each current range. You can also read the input resistance of the volts function directly with a second meter on ohms, with care.

Why the two internal resistances matter

An ideal voltmeter would draw no current and an ideal ammeter would drop no voltage. Real meters fall short of both. A voltmeter connected across a high-resistance circuit forms a voltage divider with it and reads low. An ammeter in series adds resistance that reduces the very current you are measuring, a voltage drop manufacturers call burden voltage.

Digital meters usually have a voltmeter input resistance of about 10 megohms on most DC ranges, set by a precision resistor divider. Some offer a much higher impedance on the lowest millivolt range, and some cheap meters sit nearer 1 megohm. Classic analog meters are specified in ohms per volt, for example 20 kilohms per volt, so on a 10 V range the input resistance is 200 kilohms, which changes with every range. Ammeter resistance comes from a shunt resistor plus the fuse, the jack contacts and the test leads. It is lowest on the high-current range and highest on the microamp range.

Typical values to expect

Meter and function Typical internal resistance Where it comes from
Handheld DMM, DC or AC volts About 10 megohms Precision input divider
Some DMMs, lowest mV range Can be much higher, sometimes gigaohms Direct input to the converter
Budget DMM, volts Around 1 to 10 megohms Varies by design; check manual
Analog meter at 20 kilohms per volt, 10 V range 200 kilohms Multiplier resistor plus movement
DMM, 10 A range Tens of milliohms or less, plus fuse and leads Low-value shunt
DMM, mA range A fraction of an ohm to a few ohms Larger shunt, fuse resistance
DMM, microamp range Hundreds of ohms to kilohms Large shunt to give a measurable drop

The manual often states the input impedance for volts and the burden voltage, for example a figure in millivolts per milliamp or millivolts per amp, for current ranges. Measuring them yourself is a good check on a used or unknown meter.

Measuring voltmeter input resistance

Method 1: series resistor divider. This is the cleanest method because the meter is measuring itself.

  1. Choose a steady DC source, such as a fresh 9 V battery or a bench supply.
  2. Measure the source voltage directly with the meter: call it Vs.
  3. Pick a known resistor close to the expected input resistance, for example 10 megohms. Measure its actual value on the ohms range of a second meter if you can, because high-value resistors often have a tolerance of a few percent.
  4. Connect the resistor in series between the source positive and the meter’s V input. Connect the meter’s COM to the source negative.
  5. Read the meter: call it Vm.
  6. Calculate input resistance: Rin = R times Vm divided by (Vs minus Vm).

Worked example. The battery reads 9.40 V directly. With a 10.0 megohm resistor in series, the meter reads 4.70 V. Vs minus Vm is 4.70 V, so Rin = 10.0 times 4.70 divided by 4.70 = 10.0 megohms. If it had read 6.27 V instead, Rin = 10.0 times 6.27 divided by 3.13, about 20 megohms. Keep your fingers off the resistor and leads during the reading, because your skin resistance in parallel will skew a measurement at these values.

Method 2: a second meter on ohms. Switch the meter under test to DC volts and leave it on. Put a second meter on its megohm range across the first meter’s V and COM jacks. The second meter injects a tiny test current and reads roughly the input resistance. This is quick but less certain, since some meters’ input protection or auto-range circuits behave unpredictably with an external ohmmeter attached.

Measuring ammeter resistance and burden voltage

  1. Set up a simple current source: a battery and a resistor that limits current to well within the range you are testing. For a 200 mA range, a 9 V battery and a 100 ohm resistor gives roughly 90 mA.
  2. Put the meter under test in series, on the range you want to characterise, with the red lead in the correct mA or A jack.
  3. Read the current: call it I.
  4. With a second meter on DC millivolts, measure directly across the first meter’s input jacks, on the jack metal rather than at the probe tips, to exclude the test leads. Call it Vb.
  5. Calculate the ammeter resistance: Ra = Vb divided by I.

Worked example. The meter reads 90.0 mA and the second meter shows 45 mV across its jacks. Ra = 0.045 V divided by 0.090 A = 0.5 ohm. Repeat with the probes included in the measurement and you will usually see the leads add a small fraction of an ohm more. On the 10 A range with 1.00 A flowing and 30 mV measured, Ra = 0.03 ohm. Bench meters such as the Labloot LB1041 bench multimeter are convenient for the second measurement because of their extra resolution.

Do not try to measure ammeter resistance by putting the meter’s current jacks directly across a battery. That is a short circuit through a fraction of an ohm and will blow the fuse or worse.

Pitfalls and what the numbers mean in practice

  • Ranges differ. Measure each volt and current range you care about. The microamp and milliamp shunts are very different from the 10 A shunt.
  • The fuse is part of the resistance. A partly damaged fuse raises ammeter resistance. If one meter shows much higher burden than its manual suggests, check the fuse.
  • Autoranging can switch mid-test. Lock the range manually where possible so you know which shunt or divider you are measuring.
  • High-value resistors and humidity. Surface leakage on circuit boards and your fingers can add parallel paths at megohm levels. Work on a clean, dry surface.
  • Loading error. If you measure across a 1 megohm resistor in a divider with a 10 megohm meter, the parallel combination is about 0.91 megohm, so the meter reads noticeably low. That same loading explains why a disconnected wire can show a phantom voltage on a high-impedance meter, described in why ghost voltage appears on a multimeter.

For choosing a meter with low burden voltage or high input impedance for electronics work, our guide on how to choose a digital multimeter explains which specs to look for. A general-purpose meter such as the VEVOR 6000-count multimeter is fine for these experiments as the meter under test.

Run the divider test with a 10 megohm resistor

Measure a 9 V battery directly, then again through a 10 megohm resistor in series, and plug both readings into Rin = R times Vm divided by (Vs minus Vm). If you get close to 10 megohms, your meter matches the usual DMM design, and you now know exactly how much it loads high-resistance circuits.

Frequently Asked Questions

Why is DMM input resistance usually 10 megohms?

It is a practical balance. Higher values make the meter more prone to noise and ghost voltages; lower values load circuits more. A precision 10 megohm divider is easy to build accurately and became the common standard.

Can I measure the input resistance with the meter switched off?

Not reliably. With the meter off, the input circuit may be disconnected or behave differently. Measure it on the function and range you actually use.

Why does my meter read low on a high-resistance circuit?

The meter's input resistance sits in parallel with part of the circuit and pulls the voltage down. The error grows as the circuit's resistance approaches the meter's 10 megohms.

What is burden voltage on a multimeter?

It is the voltage drop across the meter's current shunt, fuse and connections when measuring current. It reduces the current in the circuit slightly, which matters most in low-voltage circuits.

Does a better meter have lower ammeter resistance?

Often, but not always. Some meters trade a higher shunt resistance for better resolution on small currents. Check the stated burden voltage for the range you need.