How to Calibrate a Megger Tester

Your insulation tester, often called a megger, is due for a check, or its readings no longer match a colleague's. Here is what you can verify yourself, what only a lab can do, and how to stay safe with the test voltages involved.

As an Amazon Associate, Akermin earns from qualifying purchases. Product links on this page are affiliate links.

Formal calibration of an insulation tester is done by the manufacturer or an accredited lab, which checks its resistance ranges, test voltages and other functions against traceable standards and issues a certificate, usually on a yearly cycle. Between those visits you verify it yourself: with the leads apart it should read above its maximum range, with them shorted it should read close to zero, and on a known high value resistor rated for the test voltage, or a purpose-made check box, it should read within its stated accuracy. If any check fails, stop using it for sign-off work and send it for calibration or repair.

What an insulation tester measures, and what drifts

Megger is a brand name that became the everyday word for any insulation resistance tester. These instruments apply a high DC voltage, commonly a few hundred to a thousand volts or more, between a conductor and earth or between conductors, measure the tiny leakage current, and display the resistance in megohms or gigohms.

Worked example: at a 500 V test, a leakage current of 1 microamp means a resistance of 500 / 0.000001, which is 500 megohms. At 10 microamps it is 50 megohms. The measurement depends on two things being right: the actual test voltage, and the accuracy of the current measurement. Both can drift with age, damage or a failing battery, which is what calibration checks.

Multifunction installation testers add continuity, loop impedance and RCD tests, each with its own calibration points. The difference between an insulation test and a simple continuity check is covered in our guide to insulation testing versus continuity testing.

Field checks you can do before each job

Check How Expected result Failure means
Battery Use the battery check or watch for a warning Healthy level Low test voltage and false readings; replace first
Lead inspection Look for cuts, cracked insulation, loose probe tips No damage Safety risk and leakage; replace leads
Open circuit Leads apart, run an insulation test Above the top of the range Leakage in leads or meter; investigate
Short circuit Leads clipped together, run an insulation test Close to zero Lead or connection fault
Continuity null Leads shorted on the low ohms range, null if available Close to zero after null Worn leads or connectors
Known resistor Test a high voltage resistor or check box Within stated accuracy Calibration drift; send for calibration

The open and short checks are quick confidence tests, not calibration. They prove the leads are intact and the meter responds at both ends of its scale, but they do not prove that a 50 megohm reading is truly 50 megohms.

Verifying accuracy with known resistors

  1. Get a calibration check box made for insulation testers, or high value resistors that are specifically rated for the test voltage. An ordinary small resistor may be rated for only a few hundred volts and can arc across or change value at 1,000 V, giving a false result.
  2. Choose values across the range you use, for example 1 megohm, 10 megohms and 100 megohms, and know their actual values, ideally from a calibration certificate.
  3. Keep the resistors clean and dry. Fingerprints and moisture on the body create leakage paths that lower the reading at high resistance values.
  4. Connect the tester to one resistor, select the test voltage, and run the test.
  5. Record the reading and compare it with the resistor value and the tester’s stated accuracy for that range.
  6. Repeat at each test voltage you use, since the meter may behave differently at 250 V, 500 V and 1,000 V.
  7. Discharge each resistor fully before handling it; most testers do this automatically at the end of a test, but wait for the display to confirm.

Worked example: a 10 megohm check resistor reads 10.3 megohms on a tester specified at plus or minus 3 percent plus 2 digits on that range. Three percent of 10 is 0.3, plus two digits of 0.01, gives an allowance of 0.32 megohms, so 10.3 passes.

For the low resistance side, a milliohm meter or a known low value standard verifies the continuity function; our page on milliohm meters for low resistance testing explains when a four-wire instrument such as the Extech 380560 milliohm meter is the better tool.

Checking the test voltage

A tester that applies too low a voltage will report higher resistance than reality on some insulation, and one that applies too high a voltage can stress sensitive equipment. Labs measure the output voltage under load. In the field, some testers display the actual voltage during the test, which is a useful indicator.

Measuring the output with an ordinary multimeter is not straightforward: the meter’s own input resistance loads the tester, and the voltage may exceed the multimeter’s rating. Leave voltage verification to a lab unless you have a high voltage probe and the training to use it.

A practical clue comes from your known resistor checks. If readings are accurate at low resistance values but drift progressively at the highest values, or change noticeably between a fresh and a part-used battery, the output voltage or the current measurement circuit may be weakening. Note which ranges and voltages are affected when you send the tester in; it helps the lab diagnose the fault and tells you which past results might need repeating.

Lab calibration and safety

Send the tester for calibration on the interval your company, scheme or regulator sets, commonly once a year, and also after a drop, exposure to water, a failed field check, or if it was accidentally connected to a live circuit. A lab checks each resistance range and test voltage, the continuity and other functions, and the voltage warning circuit, then provides a certificate listing errors before and after any adjustment. Multifunction testers like the Kyoritsu KEW 6010B have several functions that each need checking. Our roundup of insulation resistance testers notes how models handle calibration and verification.

Insulation testers produce voltages that can give a painful or dangerous shock. Never test a circuit that is live; isolate, lock off and prove dead first with an approved voltage indicator. Disconnect sensitive electronics that could be damaged by the test voltage. Keep hands off the circuit and the probes during the test, and let the tester discharge the circuit before you touch anything, especially long cables and capacitive loads that store charge. If you are not trained and authorised for this work, it belongs to a qualified electrician.

Build your verification routine

Put together a small kit of a voltage-rated check box or resistors with known values, and log open, short and known resistor readings each month or before important sign-offs. Keep the log with the lab certificate; when a reading drifts outside tolerance, you will see it early and have the evidence to justify sending the tester in.

Frequently Asked Questions

How often should a megger be calibrated?

Yearly lab calibration is the common practice, with user verification checks between visits. Your company, insurer or certification scheme may set a specific interval, and a drop or failed check means calibrate sooner.

Can I use ordinary resistors to check an insulation tester?

Only if they are rated for the test voltage. Standard small resistors may be rated for a few hundred volts and can break down at higher test voltages, which gives misleading results.

Why does my insulation tester read lower on a humid day?

Moisture on cable ends, terminals and test leads creates surface leakage that lowers insulation resistance. Clean and dry the connections, and use the guard terminal if the tester has one.

What does the guard terminal do?

It diverts surface leakage current away from the measurement so the tester reads only the leakage through the insulation itself. It matters most at very high resistance values.

Is a Megger the same as a multimeter on ohms?

No. A multimeter tests resistance at a very low voltage, which cannot reveal insulation that breaks down under stress. An insulation tester applies hundreds of volts to expose those faults.