Remove the probe and terminate the channel input, with a 50 ohm terminator if the scope has a 50 ohm input setting or a shorting or terminating BNC cap on a 1 megohm input. Set the channel to its most sensitive volts per division, full bandwidth, normal sample acquisition with no averaging, and a timebase that captures plenty of points. Then use the scope’s measurement function to read AC RMS, or standard deviation, of the flat trace: that number is the noise floor at that setting. Repeat at each vertical range and with bandwidth limit on, because noise changes with both.
Where scope noise comes from
Every oscilloscope adds a little random voltage of its own. The input attenuator, the front end amplifier, the analogue to digital converter and the reference circuits all contribute. On screen it shows as a thickened, fuzzy baseline instead of a thin line. On high sensitivity ranges the amplifier’s noise dominates; on less sensitive ranges, quantisation in the ADC and attenuator noise matter more.
The noise floor sets the smallest signal you can see clearly. A ripple of 2 mV is easy on a scope with 100 µV RMS of baseline noise and hard on one with 1 mV RMS. Knowing your scope’s figure also tells you when a noisy trace is your circuit and when it is the instrument. If you are still deciding whether a scope or a meter suits a job, see oscilloscope versus digital multimeter; a multimeter averages away fast noise and will never show it.
Setting up the measurement
The goal is to measure only the scope, so nothing outside should be connected to the input.
- Warm up. Leave the scope on for at least 20 to 30 minutes so the front end temperature settles.
- Terminate the input. Disconnect the probe. On a scope with a selectable 50 ohm input, set it to 50 ohm and fit a 50 ohm BNC terminator, or leave it open, depending on what you want to characterise. On a 1 megohm only input, fit a metal shorting cap or a 50 ohm terminator. An open BNC on a 1 megohm input acts like an antenna and picks up hum.
- Vertical setting. Choose the most sensitive volts per division, for example 1 mV/div or 2 mV/div. Set the offset to zero so the trace sits in the middle.
- Bandwidth. Start with full bandwidth, bandwidth limit off.
- Acquisition. Use normal or sample mode. Turn off averaging, high resolution mode and any filtering, which all lower noise and would hide the real floor.
- Timebase and memory. Choose a timebase that gives a long record, so the statistics settle, and auto or free running trigger so it updates continually.
- Measure. Add an AC RMS measurement, also called standard deviation or RMS AC on some scopes, plus peak to peak. Turn on statistics and let them gather a few hundred acquisitions.
Use the same method on a handheld 3 in 1 scope meter such as the FNIRSI 2C53T scope meter; if it lacks an RMS AC measurement, record peak to peak and estimate RMS as described below.
Reading RMS versus peak to peak
Random noise is roughly Gaussian. The RMS value, which is the standard deviation of the trace about its mean, is stable and the figure most datasheets quote. Peak to peak grows the longer you watch, because rare large excursions keep appearing. For a long record, peak to peak is typically about six to eight times the RMS value.
| Measurement | What it tells you | Use it for |
|---|---|---|
| AC RMS or standard deviation | Average noise power, stable from capture to capture | Comparing scopes and settings |
| Peak to peak | Worst excursion seen so far | Judging trace thickness and trigger reliability |
| Mean | DC offset of the baseline | Checking zero and offset calibration, not noise |
Worked example: at 1 mV/div with full bandwidth, the scope reports 110 µV RMS and 0.82 mV peak to peak. The ratio is about 7.5, consistent with random noise. Turning on a 20 MHz bandwidth limit drops RMS to 55 µV. Since random noise scales with the square root of bandwidth, a factor of two drop means the bandwidth was reduced about four times, which matches a scope of roughly 70 to 100 MHz.
You can also express noise as a density: divide RMS by the square root of bandwidth in hertz. 110 µV over about 100 MHz is 110 µV / 10,000, roughly 11 nV per root hertz, a useful figure when comparing to a circuit’s own noise.
Probes, settings and what changes the floor
The terminated input measurement is the scope alone. Real measurements add the probe and its ground path, which often contribute far more noise than the scope.
- 10x probes multiply noise. A 10x probe divides the signal by ten before the scope. The scope’s own noise is unchanged at its input, so referred to the probe tip, it is ten times larger. 110 µV RMS at the input becomes 1.1 mV RMS at the tip. For small signals, use a 1x probe if bandwidth allows, or a dedicated low noise probe.
- Long ground leads pick up interference. The loop of a 15 cm ground clip acts as an antenna for switching noise. Use the short ground spring that came with the probe.
- Bandwidth limit helps. If your signal is slow, a 20 MHz limit can halve the noise or better.
- Higher resolution modes help. High resolution or averaging trade bandwidth or update rate for lower noise, legitimately, once you know what you are giving up.
- Volts per division matters. Measure the floor at each range you use. Some scopes show a jump in noise at the ranges where an extra internal attenuator switches in.
Checking your result and when the scope is at fault
Compare your results with the datasheet’s noise specification, making sure the conditions match: same input impedance, bandwidth and vertical setting. Some datasheets quote RMS noise at 50 ohm with bandwidth limit on, which makes the figure look smaller than a full bandwidth 1 megohm test. If your scope is within the stated figure, the noise floor is normal and the fix for a noisy measurement is in your probing or the circuit.
Suspect a fault if the noise is several times the specification at every range, if one channel is much noisier than the others under identical settings, or if you see a regular pattern, such as a fixed frequency spike, on a terminated input. Check that the scope is not sitting next to a switching power supply or phone charger, and try a different mains outlet. A persistent pattern on a terminated input, with nothing nearby, is a matter for the manufacturer’s service.
For safety, terminating the input means nothing hazardous is connected, but remember for later measurements that the probe ground on most bench scopes is tied to mains earth. Clipping it onto a mains referenced circuit can cause a short. The article on multimeter versus oscilloscope covers where each instrument is the safer choice on mains equipment.
Build your own noise table
Fit a terminator, then record AC RMS and peak to peak at every volts per division setting, once with full bandwidth and once with the 20 MHz limit; tape that table inside the lid, and next time a trace looks fuzzy you will know at a glance whether the noise is the scope or the circuit.
Frequently Asked Questions
Should I measure scope noise with the probe connected?
Measure the scope alone first with a terminated input, because that gives a repeatable figure to compare with the datasheet. Then measure with your probe tip shorted to its ground spring to see what the probe adds. Both numbers are useful.
Why does my oscilloscope show noise with nothing connected?
An open input on a 1 megohm channel picks up hum and radio interference like an antenna, and the scope's own amplifier noise is always present. Fit a terminator or shorting cap to see the true internal floor.
Is RMS or peak to peak noise more useful?
RMS is the stable figure for comparing instruments and settings, which is why datasheets use it. Peak to peak describes how thick the trace looks and keeps growing with capture time.
Does averaging reduce the noise floor?
Averaging reduces random noise on repetitive signals, roughly by the square root of the number of averages. It does not help on single events, and it hides the real noise floor, so turn it off when measuring the floor.
Why is the noise higher with a 10x probe?
The probe attenuates the signal by ten, but the scope's own noise stays the same at its input. When the scope scales the display back up to tip voltage, the noise is scaled by ten as well.