Only some Fluke multimeters measure temperature, so first check the dial or the front panel for a temperature position or a degree symbol. If it is there, plug a K-type thermocouple such as the bead probe that comes with many models into the COM and V jacks, with the plug’s positive side in the V jack, turn the dial to temperature, and touch or immerse the probe tip on the thing you want to measure. Give it a few seconds to settle, switch between Celsius and Fahrenheit with the function button if needed, and check the setup once in stirred ice water, which should read close to 0 degrees Celsius.
Which Fluke meters can do it
Temperature is a feature of particular models, not something every multimeter does. Among Fluke’s handheld digital multimeters, the general rule is that the HVAC, industrial and advanced models include it and many basic electrician models leave it out. The 87V family, for example, includes a temperature function, and in the small HVAC range the 116 includes temperature while its sibling the 117 does not. If you are unsure, look for:
- A dial position marked with a thermometer symbol or with degrees Celsius or Fahrenheit.
- A thermocouple probe supplied in the box, usually a black bead-type probe with a two-pin plug.
- Temperature listed in the specification sheet or the manual for your model.
If your meter has none of these, it cannot read temperature, even with a probe plugged in. A thermocouple produces only a few hundredths of a volt, and the meter needs a dedicated mode with cold junction compensation to turn that into degrees. If you are choosing a meter and want this feature, the guide on how to choose a digital multimeter covers which extra functions are worth paying attention to. Bench meters such as the OWON XDM1241 desktop multimeter also offer temperature inputs if you work mainly at a desk.
How a thermocouple reading works
A K-type thermocouple is two wires of different alloys, nickel-chromium and nickel-aluminium, joined at the tip. When the tip is at a different temperature from the other end, where the wires connect to the meter, a small voltage appears, roughly 41 microvolts per degree Celsius for type K over its common range. The meter measures that voltage, measures the temperature of its own input terminals with an internal sensor, and adds the two together. That internal step is called cold junction compensation, and it is why the meter needs a moment to settle if it has just come in from a cold van.
| Item | Typical value or note |
|---|---|
| Probe type for Fluke handhelds | K-type thermocouple |
| Signal size | About 41 microvolts per degree Celsius |
| Bead probe supplied with many models | Rated roughly from minus 40 to 260 degrees Celsius |
| Standard K-type wire accuracy | Around 2 degrees Celsius, or a small percentage at higher temperatures |
| Display with no probe | Over-range or open indication |
The probe’s temperature rating matters more than the meter’s range. The meter may accept a reading far above what the little bead probe can survive. For ovens, exhaust or flame, use an immersion or surface probe rated for that temperature.
Taking the measurement, step by step
- Disconnect from any live circuit. The thermocouple shares the voltage input jacks. Do not leave test leads in a live circuit while using the temperature probe, and never clamp a bare thermocouple onto a live conductor.
- Plug in the probe. The common Fluke bead probe has a plug sized for the COM and V jacks. One side is marked positive or has a polarity key; that side goes into the V jack, the other into COM. Probes with a miniature thermocouple connector need an adapter that converts to banana plugs.
- Select temperature. Turn the dial to the temperature position. On some models temperature shares a dial position with another function, and you press the function button to step to it.
- Choose the unit. Switch between Celsius and Fahrenheit using the method your manual describes, often a button press in temperature mode or a setting held at power-up.
- Let it settle. With the probe in air, the reading should be close to room temperature after a few seconds.
- Make contact. For a liquid, immerse the tip and stir. For a pipe or surface, press the bead firmly against clean metal, and cover it with a little foam insulation or tape so air does not cool it. For air, hold the probe in the stream, away from your hand.
- Read the stable value. Wait until the last digit stops changing by more than one or two counts. Use MIN MAX on meters that have it to record extremes over time.
Checking your meter and probe in ice water
The quickest proof that your setup reads correctly is an ice bath, which is a far better reference than room air or a kitchen thermometer.
- Fill a cup to the brim with crushed ice and top it up with cold water so the gaps are filled.
- Stir for a minute.
- Put the probe tip into the middle of the ice, not touching the cup wall, and stir gently.
- Read after it settles. Within about 2 degrees Celsius of zero is normal for a standard K-type probe and meter combination.
If it is further off, check the polarity of the plug, the cleanliness of the jacks and the condition of the probe lead. A probe with a kinked or damaged wire near the tip can read badly or intermittently, and replacing the probe usually solves it. If two different probes read the same error, the meter’s temperature function may need service.
Readings that wander, lag or look wrong
| Problem | Likely cause | Fix |
|---|---|---|
| Over-range or open shown | Probe not plugged in fully, broken probe wire | Reseat the plug; try another probe |
| Reading goes down when it should go up | Plug reversed | Swap the plug round so positive is in the V jack |
| Slow to settle after moving indoors | Meter terminals still adjusting temperature | Give it several minutes at room temperature |
| Surface reading too low | Poor contact, air cooling the bead | Press firmly, insulate the tip |
| Number jumps around near motors or drives | Electrical noise on the probe lead | Route the lead away; see the guide below |
Unstable readings are covered more broadly in why multimeter readings jump around. If you want a meter for home use that includes temperature, the roundup of multimeters for home and DIY notes which models have it. For electrical and industrial work where you also need a high CAT rating, the Fluke 87V Max, rated CAT III 1000 V and CAT IV 600 V, is the kind of meter to look at; confirm the temperature function and the probe supplied on the spec sheet before you buy.
Try it now on something you know
Before you rely on the temperature mode for a job, plug in the bead probe, make an ice bath and confirm it reads close to zero, then put it in a mug of freshly boiled water and watch it climb toward boiling. If both look sensible, write the ice bath error on a sticker inside the meter case so you know the correction for that probe.
Frequently Asked Questions
Why does my Fluke meter show OL in temperature mode?
That over-range or open indication means the meter does not see a working thermocouple. Push the plug fully home, check it is a K-type probe, and try a different probe if it persists.
Can I use any K-type thermocouple with my Fluke meter?
Yes, as long as it is type K and connected through the correct plug or adapter with the polarity right. Choose the probe style and temperature rating for the job.
Is an infrared thermometer better than the multimeter probe?
They do different jobs. A thermocouple in contact measures the actual surface or liquid temperature, while an infrared thermometer reads emitted radiation and can be badly wrong on shiny metal.
Why does the reading change when I hold the probe?
Your hand warms the probe and lead. Hold the cable further from the tip and let the bead reach the temperature of what you are measuring.
Can I measure temperature on a live electrical part?
Do not touch bare thermocouples to live conductors, because the probe connects directly to the meter inputs. Isolate the circuit first or use a non-contact method designed for live equipment.