Start by proving the fault: set the meter to its lowest ohms range, touch the probe tips together, and flex each lead along its length, especially the first few centimetres at each end. A good pair reads well under one ohm and stays steady; a reading that jumps to OL or wanders while you bend the cable shows exactly where the break is. Loose banana plugs and dirty tips can often be cleaned or tightened, but a lead with a broken conductor, cracked insulation or a damaged probe body should be replaced, not spliced, if you will ever use it on mains voltage.
Why test leads fail where they do
A test lead is a fine-stranded copper wire inside thick, flexible insulation, with a banana plug at one end and a probe at the other. The copper is finely stranded so the lead stays supple, but every bend works those strands a little. The cable is stiffest where it enters the plug and the probe handle, so all the flexing concentrates there, and that is where strands fatigue and snap one by one.
For a while the remaining strands carry the signal and nothing seems wrong. Then only a few are left, the resistance rises, and the connection becomes intermittent: perfect when the lead lies straight, open when it bends. Because the insulation is intact, the break is invisible. Many people decide the meter is faulty, or that the circuit is behaving strangely, when the real cause is a half-broken lead. If your readings flicker for no reason, it is worth ruling the leads out before reading about why multimeter readings jump around for other causes.
The other wear points are the banana plug, whose sprung contacts lose tension after years of insertion, and the probe tip, which can oxidise, bend, or come loose in its holder on leads with replaceable tips.
Symptoms and what each one points to
| Symptom | Likely cause | Fix |
|---|---|---|
| OL with probes touching | Fully broken conductor, or a blown fuse if you are on a current range | Check the dial position first, then flex test; replace a broken lead |
| Reading flickers when the cable moves | Partly broken strands near plug or probe | Replace the lead; a temporary repair is for low-voltage use only |
| Probes touching read 2 to 10 ohms | Corroded tip, worn plug, or few remaining strands | Clean tips and plugs, then retest; replace if it stays high |
| Continuity beep stutters | Loose banana plug in the jack | Reseat the plug, gently restore spring tension |
| Tip wobbles or falls out | Worn threaded or press-fit tip holder | Fit a replacement tip of the same type, or replace the probe |
| Cracked, cut or melted insulation | Mechanical or heat damage | Retire the lead immediately; no repair |
One check saves a lot of confusion. If both leads look fine but the meter reads OL on every function, the problem may be the meter’s input or a fuse, not the leads. Swapping in a second known-good pair settles it in seconds.
Pinpointing the break, step by step
- Disconnect from any circuit. Take the leads off whatever you were measuring and make sure the meter is in the COM and V-ohm jacks, not the amps jack.
- Select the lowest resistance range or the continuity beeper. If you are unsure how continuity works, our guide to what continuity is and why meters beep explains the threshold.
- Touch the tips together firmly. Note the figure. A healthy pair usually shows a few tenths of an ohm.
- Test each lead separately. Plug only the black lead into COM and the red into the V-ohm jack, then clip or press the two probe tips together while you flex one lead at a time. Bend it hard near the plug, then along the middle, then near the probe handle.
- Watch the display, or listen. The point where the reading jumps or the beep cuts out is the break.
- Check the plugs. Wiggle each banana plug in its jack. If the reading changes, the plug contacts have lost tension or the jack is worn.
- Inspect the tips. Dull grey or green tips, or tips with residue from flux or battery terminals, raise contact resistance.
What you can repair, and what you should not
Safe, simple fixes. Clean dirty tips with a little isopropyl alcohol on a cloth or a fine abrasive pad, then wipe. Reseat banana plugs fully. On plugs with a split sprung pin, spreading the halves very slightly restores grip, but do this gently because the metal can snap. Replaceable tips on better probes unscrew; fit the same type.
Repairs that compromise safety. Cutting out a broken section and soldering the lead back together, or opening a moulded probe to resolder the wire, can make a lead work again. It also removes the double insulation the lead was rated with. Tape and heat shrink do not restore the creepage distances and insulation thickness a rated lead depends on. Leads are part of the meter’s safety system; the CAT rating printed on them only holds while they are intact. A repaired lead is acceptable for battery, automotive and electronics work at low voltage, provided you mark it clearly so nobody picks it up for a panel later.
Replace, do not repair, when the insulation is cracked or cut, when the probe body is chipped, when the finger guard is broken, when you can see a different colour inner layer through the outer jacket, or when the lead has been near heat or solvent. A spare set costs little compared with the risk. Kits with multiple tips and silicone cable, such as the one supplied with the FS899L multimeter and silicone test lead kit, stay more flexible than stiff PVC leads, which reduces the fatigue that causes breaks in the first place. Whatever you buy, match or exceed the CAT and voltage rating of your meter.
Confirming the leads are good again
After cleaning or replacing, repeat the short test: tips together on the lowest ohms range, flex each lead, wiggle each plug. The reading should stay steady and low the whole time. Many meters have a relative or zero function; note the lead resistance, because on low-value measurements, such as a motor winding or a short piece of wire, a few tenths of an ohm is a meaningful part of the reading.
Then check against something known. A fresh alkaline AA cell should read a little above its nominal 1.5 volts on the DC volts range, and a resistor with a printed value should read close to it within the resistor’s tolerance band. If voltage reads correctly but resistance does not, suspect the meter’s ohms circuit or battery rather than the leads.
Finally, check the current jack path if you use it. With the red lead moved to the amps jack and the dial on a current range, a blown fuse produces OL or zero no matter how good the leads are. Our guide on choosing a digital multimeter covers fuse ratings and why they matter.
Habits that make leads last
- Coil leads loosely around your hand rather than wrapping them tight around the meter body, which kinks them at the plug.
- Pull plugs out by the plug, never by the cable.
- Keep tips capped in the bag so they do not puncture the insulation of the other lead.
- Keep leads away from soldering irons, hot engine parts and solvents.
- Inspect before every mains job: look along the full length and at both ends for cracks, cuts, or exposed metal.
Professionals often keep one pair for mains work that never goes near a car or a bench, and a cheaper pair with clips for everything else. That separation keeps your safety-rated pair in rated condition.
Your two-minute next step
Put the meter on its lowest ohms range now, join the tips, and flex each lead from end to end while watching the display; if any spot makes the reading jump, retire that lead for mains work today and order a replacement pair with a CAT rating that matches your meter.
Frequently Asked Questions
What should multimeter leads read when shorted together?
Usually somewhere around 0.1 to 0.5 ohms on the lowest range, depending on lead length, plug condition and the meter. The exact figure matters less than stability. A value that wanders, or that is several ohms, points to a worn plug, a dirty tip or broken strands.
Can I use any test leads with my multimeter?
Physically most meters accept standard 4 mm shrouded banana plugs, but the leads must carry a CAT and voltage rating at least equal to the meter's and to the circuit you are testing. A meter rated for CAT III is only as safe as the weakest lead plugged into it. Cheap unrated leads are fine for low-voltage electronics only.
Why does my meter beep on continuity only sometimes?
An intermittent beep with the probes pressed together is almost always a lead or plug problem. Flex each lead and wiggle each plug to find it. If the beep is solid with the probes together but stutters on the circuit, the problem is in the circuit or your contact pressure.
Is it safe to solder a broken test lead?
It works electrically, but it removes the lead's rated insulation, so treat the result as a low-voltage lead only. Never use a repaired lead on household mains or in a distribution panel. Mark it so it is not mistaken for a rated lead.
How do I know if it is the leads or the meter?
Swap in a second pair of known-good leads. If the fault follows the old leads, they are the problem. If the fault stays with the meter, check its battery, the dial position, the input jacks and the fuses.