On most digital oscilloscopes, press the Horizontal Menu or Display button and change the time base or format setting from YT to XY; on analog scopes, turn the timebase knob to its X-Y position or press the X-Y button. Channel 1 then drives the horizontal axis and channel 2 the vertical axis, so you need both channels connected, with matching coupling and similar volts-per-division settings. Start with both channels in GND coupling, centre the dot, then switch to DC coupling and adjust the scales until the figure fits the screen.
What changes when the scope goes from YT to XY
In normal YT mode, time sweeps the trace from left to right and the input voltage moves it up and down. In XY mode, the time sweep is switched off for display purposes. The voltage on one channel, almost always channel 1, moves the spot left and right, and the voltage on the other, channel 2, moves it up and down. What you see is the relationship between the two signals, not either one against time.
Two identical sine waves in phase draw a diagonal straight line. Shift one by 90 degrees and the line opens into a circle or ellipse. Use different frequencies with a simple ratio and you get the classic Lissajous loops. With a test circuit, XY mode can also show a component’s current against voltage, which is the basis of the curve tracer view used to check diodes and capacitors.
The trigger, the time measurements and most automatic readouts stop meaning much in XY, because there is no time axis on the screen. Amplitude cursors still work and are useful for the phase method below. Many digital scopes also keep the normal YT traces available in a split view, which helps you confirm each signal is present and sized correctly before you study the XY figure.
If you are deciding whether a scope or a meter suits a job, our comparison of an oscilloscope versus a digital multimeter explains why only the scope can show relationships like this.
Where to find the setting
Menu names vary between brands and firmware versions, but XY is nearly always in one of a few places.
| Scope type | Where to look | Typical label |
|---|---|---|
| Bench digital scope | Horizontal Menu, Time Base softkey | YT, XY, Roll |
| Bench digital scope, other layouts | Display button, Format softkey | YT or XY |
| Some newer digital scopes | Acquire menu or a dedicated XY item in the trace menu | XY Display On |
| Analog scope | Timebase knob turned fully to the end of its range | X-Y |
| Analog scope, alternative | Front panel push button near the timebase | X-Y or CH1 X |
| Handheld or combination scope | Main menu or display settings, if supported | XY, or not available |
XY needs two input channels. A two-channel handheld such as the FNIRSI 2C53T 2CH lists XY mode among its features, but not every compact scope implements an XY display, so check the menu or manual of yours. If you cannot find it after checking these menus, search the manual for “XY” or “Lissajous”.
Getting a clean figure, step by step
- Connect the signals. The signal you want on the horizontal axis goes to channel 1, the vertical one to channel 2. Both probes’ ground clips go to the same ground point.
- Set the probe ratio in each channel menu to match the probes, normally 10x for both.
- Set both channels to GND coupling and switch to XY. You should see a single dot. On an analog scope, turn the intensity down first so a bright stationary dot does not burn the phosphor.
- Centre the dot using the position controls: channel 1 position moves it sideways, channel 2 position moves it up and down.
- Switch both channels to DC coupling, or both to AC if the signals ride on a DC level you want to ignore. Keep them the same so you do not introduce a phase error.
- Match the scales. Set both channels to the same volts per division if the signals are similar in size, then adjust so the figure fills most of the screen without clipping.
- On a digital scope, set the timebase so the acquisition window holds several full cycles of the slower signal. The timebase still controls how much data is captured even though it no longer appears as time.
If you only get a vertical or horizontal line, one channel has no signal or is still on GND. If you get a dot, both are missing or grounded. A shape that slowly rotates means the two frequencies are close to, but not exactly at, a simple ratio.
Reading phase and frequency from the shape
Phase between two signals of the same frequency. Centre the ellipse. Measure the distance from the centre to where the ellipse crosses the vertical axis, call it A, and the distance from the centre to the ellipse’s highest point, call it B. Then sin of the phase angle equals A divided by B.
Worked example: the ellipse crosses the vertical centre line 2 divisions above centre, and its top is 4 divisions above centre. A divided by B is 0.5, and the angle whose sine is 0.5 is 30 degrees. A straight diagonal line means 0 or 180 degrees, depending on which way it slopes; a circle with equal scales means 90 degrees. The method cannot tell you which signal leads, only by how much.
Frequency ratio. Count how many times the figure touches a horizontal line drawn along its top edge, and how many times it touches a vertical line along its side. The vertical-channel frequency divided by the horizontal-channel frequency equals the top touches divided by the side touches. A figure eight lying with two lobes along the top and one point on the side means channel 2 runs at twice the frequency of channel 1.
For the strengths and limits of each instrument on measurements like these, see multimeter versus oscilloscope.
Errors that distort the figure
- Mismatched coupling. AC coupling adds a small phase shift at low frequencies. If one channel is AC and the other DC, a pair of identical signals shows as a thin ellipse instead of a line.
- Uncompensated probes. A poorly compensated probe shifts phase at higher frequencies. Compensate both probes on the scope’s calibration output before precise work.
- Frequency beyond the scope’s comfort zone. Near the bandwidth limit, the channels may not track each other perfectly, so phase readings drift.
- Low sample rate on digital scopes. If the timebase is too slow, the figure becomes jagged dots. Speed it up until the curve is smooth.
- Common ground problems. Both probe grounds connect to the scope’s earth. Do not clip them to two different points that are at different voltages, or you short those points together through the scope.
Try it on the calibration output first
Before tackling your real circuit, put both probes on the scope’s calibration square wave, switch to XY and confirm you see a diagonal pattern of dots or a short line. That proves the menu, the channel order and the probe settings, and any odd shape you see afterwards belongs to your signals, not your setup.
Frequently Asked Questions
Which channel is X in XY mode?
On nearly every scope, channel 1 drives the horizontal X axis and channel 2 drives the vertical Y axis. A few four-channel scopes let you choose which pair to use. Check the menu or the screen labels if the figure appears rotated.
Can a single-channel oscilloscope do XY mode?
Not with two signals, since XY needs one input for each axis. Some analog scopes have an external horizontal input that can act as the X channel. Without that, you need a two-channel scope.
Why does my Lissajous figure keep spinning?
The two frequencies are close to a simple ratio but not exactly on it, so the phase between them changes continuously. The slower it rotates, the closer the ratio is. Locking both signals to a common reference stops the rotation.
Is XY mode useful for audio work?
Yes. Feeding left and right channels into X and Y shows stereo phase and width: a single diagonal line indicates mono content, and a wide cloud indicates wide stereo. Inverted polarity on one channel tilts the line the opposite way.