How Often Should a FLIR Camera Be Calibrated?

Your thermal camera still produces sharp images, but a client, auditor or insurer is asking when it was last calibrated. Here is a realistic interval for the way you actually use it, and how to check it between services.

As an Amazon Associate, Akermin earns from qualifying purchases. Product links on this page are affiliate links.

For a FLIR camera whose temperature readings go into reports, the common practice and the usual manufacturer guidance is a professional calibration about once a year. If you only use the camera to find patterns, such as a cold patch on a wall or a hot breaker, the exact number matters less and many users stretch the interval, provided simple field checks still pass. Calibrate sooner after a drop, a failed check, a firmware or lens change, or before work that may end up in a dispute or an insurance claim.

What calibration actually means for a thermal camera

A thermal camera does not see temperature. Its microbolometer detector measures infrared energy, and the camera’s software converts that energy into a temperature using a stored calibration curve plus your settings for emissivity, reflected temperature and distance. How that conversion works is explained in how a thermal camera sees heat.

Over time the detector’s response and the optics can shift slightly, and that curve becomes less accurate. A calibration at a service lab points the camera at reference blackbody sources at several known temperatures across its range, compares what the camera reports, adjusts the curve if needed, and issues a certificate listing the results. That certificate is what an auditor or client is really asking for.

Calibration is not the same as the clicking you hear while the camera runs. That click is the shutter performing a non-uniformity correction, which evens out pixel-to-pixel differences. It keeps the image clean but does not correct absolute temperature accuracy.

Matching the interval to how you use the camera

How you use it Typical interval Why
Electrical and mechanical inspections with reported temperatures About every 12 months Clients and insurers rely on the figures; certificates are often requested
Work under an ISO 9001 or similar quality system As set by your procedure, commonly yearly Auditors check that measuring equipment is calibrated and traceable
Building and energy surveys, pattern finding Every one to two years, with field checks Relative differences matter more than absolute values
Process monitoring, research, medical-adjacent screening Yearly or more often, per your standard Absolute accuracy is the whole point
Occasional home use, phone-attached imagers Rarely, if ever Service calibration may not be offered or may cost more than the device

A professional camera such as the FLIR E6 Pro or FLIR E50 fits the first rows, where a dated certificate is part of doing the job properly. A phone accessory like the FLIR One belongs to the last row; it is a fine tool for spotting draughts and leaks, but it is not intended to produce certified temperature figures.

Signs you should not wait for the yearly date

  • It was dropped or knocked hard. Optics can shift and internal components can be strained even when the image looks normal.
  • A field check fails. If a known reference reads outside the camera’s stated accuracy, which for many professional cameras is around plus or minus 2 C or 2 percent of reading, stop relying on the figures.
  • The lens was changed or serviced. Interchangeable lenses are calibrated with the camera; a new lens may need its own calibration.
  • Readings disagree with other instruments in a consistent direction across several jobs.
  • A report may be contested. If the result could end up in a legal, insurance or warranty dispute, a fresh certificate removes an easy line of attack.
  • It was stored or used outside its rated temperature or humidity range, for example left in a hot vehicle for long periods.

How to field-check a thermal camera between calibrations

You cannot calibrate a camera at home, but you can verify it. The aim is to aim it at something whose temperature you know, with an emissivity you control.

  1. Prepare a high-emissivity target. Put a strip of black electrical tape or matte black paint on a metal container. Bare metal is a poor target because it reflects its surroundings; why shiny metal misleads a thermal camera explains the trap.
  2. Make a known temperature. A well-stirred ice and water mixture sits close to 0 C. Water near boiling at sea level is close to 100 C, though it cools quickly and steam interferes. A container of warm water checked with a calibrated contact thermometer gives a midrange point.
  3. Set emissivity correctly. Black electrical tape is usually around 0.95. Set the camera to match, and set reflected temperature to room temperature. The reasons are covered in emissivity in thermal imaging.
  4. Measure from close range, square to the surface, with the spot or box fully on the taped area.
  5. Compare and record. Write the date, reference temperature, camera reading and difference in a simple log.

Worked example: a contact thermometer reads 40.0 C on the water, the camera reads 41.2 C on the tape. The difference is 1.2 C, inside a plus or minus 2 C specification, so the camera passes. If the next month it reads 43.5 C under the same conditions, the 3.5 C difference is outside spec and the camera should go for service.

A spot infrared thermometer is not a good reference, because it has the same emissivity problems as the camera. Thermal camera vs infrared thermometer covers where each fits.

What to expect from a calibration service

Calibration is normally done by the manufacturer’s service centre or an accredited lab. Ask in advance what you will receive. Useful things to request are:

  • A certificate showing readings before and after adjustment, so you know how far off it had drifted. If it was far out, review recent reports that depended on absolute temperatures.
  • Calibration points across the range you actually use, not just one midpoint.
  • Traceability to national standards, and accreditation if your clients require it.
  • Any firmware updates or repairs carried out at the same time.

Build the turnaround time into your schedule, since the camera may be away for a while. Many inspection firms book calibration during their quietest season, or keep a second camera for cover.

Look after the camera between services and the yearly visit is more likely to be a formality. Keep the lens capped and clean it only with the recommended cloth and fluid, since scratches and residue scatter infrared. Store it in its case at room temperature rather than in a vehicle. Let it warm up for several minutes after switching on before taking readings you plan to report, because the detector and optics settle as they reach operating temperature. And keep firmware current, as updates sometimes improve measurement routines. None of this replaces calibration, but it reduces the drift that calibration has to correct.

Set your schedule today

Find the date on your camera’s last calibration certificate, put the next due date in your calendar about twelve months on if you report temperatures, and set up a one-page field-check log with a taped ice bath and a warm water reference so that you have evidence of accuracy between services and an early warning if it starts to drift.

Frequently Asked Questions

Does FLIR recommend yearly calibration?

For cameras used to measure temperature, yearly calibration is the commonly given guidance and widely used practice. Check the manual or service documentation for your model, as the exact advice can vary.

Is the camera's shutter click a calibration?

No. That is a non-uniformity correction that evens out the image. It does not verify or adjust absolute temperature accuracy.

Can I calibrate a FLIR camera myself?

Not in the formal sense. You can field-check it against known temperature references, but adjustment and certification need blackbody equipment at a service lab.

Does a thermal camera need calibration if I only look for cold spots?

Much less. Pattern finding relies on relative differences, which stay useful even if absolute readings drift slightly. Periodic field checks are a sensible minimum.

What happens if my camera was out of calibration for months?

Look at the before-adjustment results on the certificate. If the error was within the tolerance your reports needed, nothing changes; if it was larger, review any recent findings that relied on exact temperatures.