How to Reset a Fluke 87 Multimeter

Your Fluke 87 is stuck on a mode, beeping, showing a warning, or reading oddly, and you are looking for the reset. Here is what actually returns it to normal.

As an Amazon Associate, Akermin earns from qualifying purchases. Product links on this page are affiliate links.

There is no dedicated reset button on a Fluke 87; turning the rotary switch to OFF and then to the function you want clears Hold, MIN MAX, relative and range settings, which is what most people mean by a reset. Power-on options, the features enabled by holding a button while switching on, such as disabled auto power-off, end on most versions when you turn the dial to OFF and start the meter normally again; removing the battery for a minute is the catch-all if anything still seems stuck. When the meter still misbehaves after that, check the battery, the leads and the two current fuses before suspecting a deeper fault. Calibration adjustments are not a user reset and should be done by a qualified calibration lab.

What resets itself and what stays

The Fluke 87 family, including the 87V and the Fluke 87V Max, keeps very little user state. The modes you can trigger with the front buttons, such as HOLD, MIN MAX recording, the relative or delta reading, manual range, Peak capture and the high-resolution digit mode, are temporary. They live only while the meter stays on the same function. Turning the dial away and back, or to OFF and back, cancels all of them.

What persists while the meter stays on are the power-on options: features you enable by holding a button while turning the meter on from OFF, such as switching off the beeper, disabling the automatic power-off, or checking the display segments. On most versions they end when the dial goes back to OFF, and the meter starts normally next time unless you hold the button again. The user manual lists each option and how it is cancelled, because the exact set differs a little between versions; removing the battery is the catch-all.

Calibration constants are stored internally and are never cleared by anything you can do from the front panel. That is deliberate. A meter that could lose its calibration from a button press would be useless for compliance work.

A clean restart, in order

  1. Remove the leads from any circuit. Never troubleshoot a meter that is still connected to a live supply.
  2. Turn the dial to OFF, wait a few seconds, then select a simple function like DC volts. Hold, MIN MAX and relative modes are now gone.
  3. Check the display at start-up. All segments flash briefly on power-up on many versions. Missing segments point to a display fault.
  4. If the meter still behaves strangely, turn it off and open the battery door on the back. On the 87V the battery and fuses sit behind the rear cover, released by turning its fasteners. Remove the battery, wait a minute, and refit it, or fit a fresh one.
  5. Start normally without holding any button, so no power-on option is enabled.
  6. Run a quick functional check: short the leads on ohms, read a known battery on DC volts, and check continuity beeps.

If you see the word LEAd or hear a warning chirp, that is not a fault: the meter detected a lead in the A or mA jack while the dial is on a voltage or other non-current function. Move the red lead back to the V jack.

Problems a restart will not fix

Symptom Likely cause What to do
Reads zero or OL on current ranges only Blown current fuse Test the fuse through the meter, replace with the exact Fluke-specified fuse
Low battery icon, readings unstable Weak 9 V battery Replace the battery; low voltage degrades accuracy
OL with probes shorted on ohms Broken test lead Flex test the leads, swap in a spare pair
Turns off after a while Normal auto power-off Press a button or turn the dial to wake it; disable at power-on if needed
Readings frozen Hold or auto-hold active Press HOLD again or turn the dial
Reading jumps by large steps Manual range or low resolution mode active Turn the dial to cancel manual range
Display dead with a fresh battery Internal fault, damage, moisture Send to an authorised service centre

Checking the fuses without opening the case

The 87 series has two current fuses: one for the mA and microamp input and one for the 10 A input. Fluke describes a check using the meter itself. Set the dial to ohms, put the red lead in the V input jack, and touch the probe tip into the mA jack, then into the A jack. A good 10 A path reads a fraction of an ohm. The mA path reads much higher, on the order of a kilohm on many versions, because of the internal resistance in that input, so compare it with the good-fuse figure in your manual rather than expecting near zero. OL on either means the fuse in that path is open.

Replace a blown fuse only with the exact type and rating Fluke specifies, which are high-energy fuses designed to clear safely if the meter is ever connected across a high-energy circuit. A cheap glass fuse of the same amp rating looks similar but cannot safely interrupt that kind of fault, and fitting one removes a key part of the meter’s CAT protection. If fuses blow repeatedly, the usual cause is putting the leads across a voltage while in the current jacks, a mistake also covered in choosing a multimeter for home electrical work.

Confirming the meter reads correctly after the reset

  • Ohms: short the leads and expect a few tenths of an ohm, then measure a marked resistor and compare with its tolerance.
  • DC volts: a fresh alkaline AA should read a little above 1.5 V; compare with a second meter if you have one.
  • AC volts: a household outlet should read close to your nominal supply. Because the 87 is true-RMS, it reads correctly on distorted waveforms too; this page on true-RMS explains why that matters.
  • Continuity: the beeper should sound instantly with the leads shorted.
  • Current: measure a small known load, such as a lamp on a battery, to confirm the mA path works.

If every function checks out against known values, the meter is fine and nothing deeper needs resetting.

Calibration and service: when a user reset is not enough

If readings are consistently off compared with a trusted reference, the meter needs calibration, not a reset. Calibration adjustment on a Fluke 87 requires reference standards and a defined procedure, and is done by Fluke service or an accredited calibration lab that issues a certificate. For professional or compliance work, many organisations recalibrate on a regular schedule, commonly once a year. Send it for service, too, if the case is cracked, the input jacks are damaged, the meter got wet beyond its rating, or it was exposed to a voltage spike and now reads wrong. A damaged meter can display normal-looking numbers while its protection is compromised, which is a real safety risk on mains and industrial circuits.

A practical rule: if a meter has been dropped hard, has measured a lightning-affected or surge-damaged circuit, or was connected across a high voltage on the wrong function, retire it from live work until it has been checked, even if it seems to read normally.

The quickest fix to try right now

Turn the dial to OFF, take the leads out of every jack, swap in a fresh 9 V battery, then start the meter normally and check it on a battery and a shorted lead; if it passes both, the meter is reset and ready.

Frequently Asked Questions

Does the Fluke 87 have a factory reset?

Not in the sense of a menu or button combination that restores everything. Turning the dial to OFF clears temporary modes and, on most versions, ends power-on options; removing the battery is the catch-all. Calibration values are protected and are never erased by the user.

Why does my Fluke 87 keep turning itself off?

That is the automatic power-off, which saves the battery when the meter has been idle for a while. Turning the dial or pressing a button wakes it. The manual explains a power-on option that disables it for long recordings.

What does LEAd on the display mean?

It is a warning that a test lead is in a current input jack while the dial is not set to a current function. Move the red lead to the V jack before measuring voltage to avoid blowing a fuse or worse.

Can a blown fuse affect voltage readings?

No. The current fuses only sit in the mA and A paths, so voltage and resistance keep working with a blown fuse. That is why a meter can seem fine until you try to measure current.

Do the same steps work on the Fluke 177 or 117?

The basic approach is the same: dial to OFF clears temporary modes, a normal start without holding a button avoids start-up options, and fuse checks follow the manual. The exact buttons and fuse ratings differ, so check the manual for that model.