Turn the selector to a DCV range higher than the voltage you expect, connect black to negative and red to positive, then read the black DC scale, usually labelled DCV or with a straight line symbol. Pick the row of numbers whose right-hand end matches your range, or is a factor of ten away from it, and scale the reading: on the 50 V range read the 0 to 50 row directly; on the 500 V range read the same row and multiply by ten. Look straight down on the needle, using the mirror strip if the meter has one, so you do not read it at an angle.
How the dial and the scale relate
An analog multimeter has one moving coil movement and many ranges. The range switch puts different resistors in series with the movement so that full deflection of the needle equals the number printed next to the switch position: 2.5 V, 10 V, 50 V, 250 V, 1000 V on a typical meter. The scale plate does not print a separate arc for every range. Instead it prints a few rows of numbers on the DC arc, commonly ending in 10, 50 and 250, and expects you to use the one that matches.
So the range you select tells you what the far right of the arc is worth. Everything else is proportion. A needle halfway across is half the range, whatever the printed numbers say.
| Range selected | Scale row to read | Multiply by |
|---|---|---|
| 0.25 V | 0 to 250 | 0.001 (reading in mV divided by 1000) |
| 2.5 V | 0 to 250 | 0.01 |
| 10 V | 0 to 10 | 1 |
| 50 V | 0 to 50 | 1 |
| 250 V | 0 to 250 | 1 |
| 1000 V | 0 to 10 | 100 |
Your meter’s ranges may differ, so check that the printed rows have the same leading digits as your switch positions. A meter with 3, 12, 60 and 300 V ranges, for example, prints rows ending in 3, 12 and 60.
Step by step on a live circuit
- Check mechanical zero. With the leads disconnected, the needle should rest exactly on the left-hand zero. If not, turn the small screw under the pivot with a flat screwdriver until it does. This is different from the ohms zero knob, which you do not need for voltage.
- Plug in the leads. Black in COM or minus, red in the jack marked V, ohms and mA, usually labelled plus.
- Choose a range above the expected voltage. For a 12 V car battery, use 50 V, not 10 V. If you have no idea, start on the highest range and work down.
- Connect across the source. Red to positive, black to negative or ground, in parallel with what you are measuring.
- Watch the first movement. If the needle swings left against the stop, the polarity is reversed. Remove the probes immediately and swap them. A hard reverse kick can bend the pointer.
- Step down for resolution. If the needle sits in the first fifth of the scale, drop to the next lower range, provided the reading will not exceed it. Readings are most accurate in the upper half to two thirds of the arc.
- Read with your eye square to the needle. If the scale has a mirror band, move your head until the needle hides its own reflection. That removes parallax error.
Worked examples
A 9 V battery on the 10 V range. Read the 0 to 10 row. The needle sits two small divisions past the 9 mark, and each small division on that row is 0.2 V. The battery reads 9.4 V.
A car battery on the 50 V range. Read the 0 to 50 row. The needle sits a little past the second small division beyond 10, and each small division on that row is 1 V, so you can read it as about 12.5 V. That resolution is enough to tell a charged battery from a flat one, but not to see the difference between 12.5 and 12.6 V. For finer resolution you would want a 15 V or 25 V range if the meter has one.
A 120 V DC supply on the 250 V range. Read the 0 to 250 row directly. The needle sits four small divisions past the 100 mark, and each small division on that row is 5 V, so about 120 V.
A 1.5 V cell on the 2.5 V range. Read the 0 to 250 row, then divide by 100. The needle at 155 means 1.55 V.
To count divisions, take the gap between two numbered marks and divide by the number of small spaces between them. If 0 and 10 are separated by 50 spaces, each is 0.2.
Errors that catch people out
- Reading the wrong arc. The top arc on most analog meters is ohms, printed right to left and squashed at one end. DC volts is usually the second arc, linear and evenly spaced. The AC arc can be slightly offset on the low ranges, so use it only for AC.
- Meter loading. An analog meter draws current from the circuit. Its sensitivity is printed as ohms per volt, often 20,000 ohms per volt on DC. On the 10 V range that means 200 kohm of input resistance. Across a high resistance circuit, such as a transistor bias divider, that drags the voltage down and you read low. A digital meter with about 10 Mohm input disturbs the circuit far less.
- Parallax. Looking from the side can shift a reading by a whole division. Use the mirror.
- Weak ohms battery is not the issue. Voltage ranges do not use the internal battery, so a flat battery affects ohms but not DCV readings.
- Fluctuating needles. Some swing is real ripple or a loose probe. Unlike a digital display, the needle averages fast changes, which can be useful. If a digital meter shows numbers wandering while the needle is steady, see why multimeter readings jump around.
Confirming the meter reads correctly
Measure a fresh alkaline AA cell, which should sit around 1.55 to 1.6 V, and a known good 9 V battery. Better still, compare against a digital meter on the same source. Analog meters are often specified to a few percent of full scale on DC, so on the 10 V range a reading within 0.2 to 0.3 V of the digital value is normal. A bigger difference, a needle that sticks, or a pointer that will not return to zero suggests a damaged movement or a burnt range resistor.
For mains work, an older analog meter may have no CAT rating at all. Use it on batteries and electronics, and for household circuits use a meter rated at least CAT II, ideally CAT III, as explained in choosing a multimeter for home electrical work. A digital meter such as the Klein Tools MM325 gives a direct numeric DC reading with no scale to interpret.
Practise on a battery
Grab a 9 V battery and measure it on the 50 V, 10 V and 250 V ranges in turn, reading the correct row each time. Seeing the same voltage land in different places on the arc makes the range and scale relationship click faster than any diagram.
Frequently Asked Questions
What does DCV mean on a meter?
It stands for direct current volts, the steady one-direction voltage from batteries, solar panels and DC power supplies. The setting is often shown as V with a straight line or a straight and dashed line above it.
Which scale is DCV on an analog multimeter?
Usually the second arc from the top, often printed in black with the label DCV or DC. The top arc is normally ohms and is non-linear.
Why does my needle move to the left when I measure DC voltage?
The probes are reversed. Disconnect and swap red and black. Reverse deflection on a high reading can damage the pointer.
Do I need to zero an analog meter before measuring voltage?
Check the mechanical zero screw so the needle rests on zero with no input. The ohms adjust knob is only used for resistance measurements.
Is an analog multimeter accurate for DC voltage?
Reasonably, typically within a few percent of the full-scale value. Accuracy is best near full scale, so choose a range where the needle sits in the upper part of the arc.