Leak-check the pump with an unopened tube, then snap both sealed tips of a fresh tube, insert it into the pump with the arrow pointing toward the pump, and pull the exact number of strokes printed on the tube or its instructions, waiting for each stroke to finish fully. Read the concentration where the color change ends on the printed scale, at the furthest point of discoloration, and apply any temperature, humidity or pressure correction from the instructions. Use tubes and pumps from the same manufacturer, check the expiry date, and remember a tube gives a single spot reading, not continuous protection.
How a colorimetric tube works
Each tube is a sealed glass tube packed with a reagent that reacts with one target gas, or a group of related gases, and changes color. A hand pump draws a precise volume of air through it, often 100 mL per stroke on piston and bellows pumps. As the air passes, the target gas reacts and stains the reagent from the inlet end forward. The more gas in the sample, the further the stain travels, and the scale printed on the glass converts that length into ppm or percent.
Because the calibration depends on volume and flow rate, the pump and the tube are a matched system. A pump from a different maker may pull a different volume or at a different speed, which shifts the reading. That is why manufacturers specify their own pumps. Tubes also have cross-sensitivities: other gases can react with the reagent, causing high readings or changing the color. The instruction sheet in each box lists known interferents. For the difference between concentration units, see ppm versus LEL in gas detection.
Before you sample: checks that prevent bad readings
- Right tube, right range. Confirm the gas name and measuring range on the box match what you need. Many gases have several tubes with different ranges.
- Expiry and storage. Reagents degrade. Use tubes before their expiry date and store them as the box says, often in a cool place, sometimes refrigerated.
- Pump leak test. Insert an unopened tube and squeeze or pull the pump. It should stay compressed or locked, and the end-of-stroke indicator should not appear, for the time stated in the pump manual. If it relaxes, clean or replace the seals before use.
- Conditions. Note air temperature, humidity and, at altitude, atmospheric pressure. Many tubes need a correction outside a standard range.
- Safety. If you suspect toxic gas, oxygen deficiency or flammable gas, wear a continuous electronic monitor while sampling and do not enter the area just to take a tube reading.
Taking the sample, step by step
- Break both tips of the tube in the tip breaker on the pump, pointing the glass away from your face. Tubes are fragile; wear eye protection and gloves.
- Insert the tube firmly into the pump inlet with the arrow on the glass pointing toward the pump. A reversed tube gives no reading or a false one.
- Hold the tube inlet in the air you want to sample. For a confined space or tank, use the manufacturer’s extension hose and allow for its dead volume as the instructions describe.
- Pull or squeeze the pump through one full stroke, and wait until the end-of-stroke indicator shows the stroke is complete. Do not rush to the next stroke.
- Repeat for the exact number of strokes stated for that tube. Some tubes have two ranges, one for fewer strokes and one for more, so check which scale applies.
- Remove the tube and read it straight away, in good light, as some stains fade or keep developing.
Reading the result and applying corrections
Read the scale at the furthest point of the color change. If the front edge is uneven or slanted, read at the average of the longest and shortest points, or at the point the instructions specify. Then correct the reading if conditions require it.
| Situation | What it means | Action |
|---|---|---|
| No color change | Gas below the lower limit, or wrong tube | Record as below range; consider more strokes if the instructions allow |
| Stain beyond the highest mark | Concentration above the scale | Use a higher range tube or fewer strokes |
| Unexpected color | Possible interfering gas | Check the interferent list; confirm with another method |
| Faint, diffuse stain | Low concentration or degraded reagent | Check expiry; repeat with a fresh tube |
| Hot, cold or humid conditions | Reaction rate and volume are affected | Apply the temperature and humidity correction table |
| High altitude | Lower pressure, so each fixed-volume stroke holds fewer gas molecules and the stain reads low | Multiply by the pressure correction factor given |
Typical accuracy of detector tubes is more modest than a calibrated electronic sensor, often described in the range of tens of percent rather than a few percent. They are excellent for identifying a gas and its approximate level, and for gases where no electronic sensor is practical.
Record each result with the tube type, the scale used, the stroke count, the location and height of the sample point, the time of day and the correction applied. Many gases are heavier or lighter than air, so a sample near the floor can differ sharply from one at head height; hydrogen sulfide and many solvent vapors collect low, while methane rises. Sample at more than one height when the gas behavior is not known.
Where tubes fit, and where they do not
Tubes are spot measurements. They tell you the concentration at one place at one moment. They do not alarm, and they cannot warn you if the atmosphere changes a minute later. Confined space entry, gas leak response and continuous worker protection require a calibrated, bump-tested electronic detector such as a multi-gas monitor; the MSA ALTAIR 4XR is an example of a portable unit covering combustible gas (LEL), oxygen, H2S and CO. The case for those is explained in what a multi-gas detector is and why gas detectors need bump tests and calibration.
Used tubes contain reacted chemicals; dispose of them following the safety data sheet and local rules, not loose in household rubbish. If any reading shows a gas near an exposure limit, or you smell gas indoors, leave the area, ventilate from outside if it is safe, and call the gas utility, emergency services or your site safety lead. Interpreting results for compliance belongs to a trained industrial hygienist or safety professional. If your readings seem inconsistent with what an electronic detector shows, the usual culprits are covered in what can cause false gas detector readings.
Practice the routine before you need it
Before your next job, take one spare tube, leak-check the pump, sample clean outdoor air with the full stroke count and confirm no color change. That one dry run proves the pump seals, your stroke timing and your reading of the scale, so the first real sample is not also your first attempt.
Frequently Asked Questions
Can I use another brand of tube in my pump?
It is not recommended. Tubes are calibrated for their own maker's pump volume and flow rate, so mixing brands can make readings wrong.
Can I reuse a gas detector tube?
No for most tubes. Once the tips are broken and the reagent has reacted or been exposed to air, the tube is used. A few multi-measurement types exist, and their instructions say so explicitly.
How accurate are colorimetric tubes?
Less accurate than a calibrated electronic sensor, often by a margin of tens of percent. They are reliable for identifying a gas and its approximate concentration when used correctly.
Why is my stain color different from the box?
Another gas may be interfering, the tube may be old, or conditions may be unusual. Check the interferent list and repeat with a fresh tube.
How many strokes should I pull?
Exactly the number printed on the tube or in its instructions for the scale you are reading. Too few or too many strokes changes the sample volume and gives a wrong result.