Most infrared laser thermometers do not have a reset button. To reset one, take the battery out for at least a minute and refit a fresh one, then go into the settings (usually the MODE or SET button) and put the emissivity back to 0.95, set the units to Celsius or Fahrenheit, and turn off any lock, hold, alarm or MAX/MIN mode. Many readings people blame on a fault are really a changed emissivity setting or the wrong unit. Afterwards, check it on a stirred ice water bath, which should read close to 0 degrees C (32 F).
What a reset can and cannot change
An infrared thermometer does not measure temperature by contact. A lens collects infrared energy radiated by the surface you aim at and focuses it on a small detector, usually a thermopile. The electronics convert that energy into a temperature using the emissivity setting, which tells the meter how efficiently the surface radiates. The laser is only a pointer to show roughly where you are aiming; it plays no part in the measurement.
That means a reset can fix settings: emissivity, units, alarms, lock mode, backlight and the stored MAX, MIN or average values. It cannot fix a dirty or scratched lens, a failing detector, or a reading that is wrong because of what you are aiming at. Shiny metal, glass and steam fool every IR thermometer, reset or not. For how emissivity works across surfaces, see what emissivity is in thermal imaging, which applies equally to spot thermometers.
Resetting step by step
- Power reset. Open the battery compartment in the handle, remove the battery and wait at least a minute. Pressing the trigger with no battery helps drain any residual charge.
- Fit a fresh battery. A weak battery causes flickering, dim displays, error messages and wrong readings before it dies. Use the type the manual specifies, often a 9 V cell or AAA batteries.
- Restore emissivity. Press MODE or SET until the E or EMS symbol appears, then use the up and down buttons to return it to 0.95, the common factory default. Some models have a fixed emissivity and no menu.
- Choose units. Many have a C/F button, or a slide switch inside the battery compartment. Some cycle units through the MODE menu.
- Clear modes. Turn off LOCK (continuous measurement), HOLD, and the high and low alarms if you do not use them. Exit MAX, MIN, DIF or AVG display modes so the screen shows the live reading.
- Let it settle. If the thermometer has been in a hot car or cold van, give it 15 to 30 minutes to reach room temperature. IR thermometers correct for their own internal temperature and read wrongly while it is changing.
Models with adjustable emissivity, such as the Etekcity dual laser, which adjusts from 0.1 to 1.0, set it and the units through a short button sequence on the instruction sheet. Simpler models like the Etekcity 774 may not offer an emissivity menu at all, so check the manual; if you have lost it, the manufacturer usually publishes it online.
What the common symptoms mean
| Symptom | Likely cause | Fix |
|---|---|---|
| Reads several degrees low on most surfaces | Emissivity set too high or too low; lens dirty | Reset emissivity to 0.95; clean lens gently |
| Reads far too low on pans or metal | Shiny surface reflects surroundings | Measure a matte spot, tape or paint mark |
| HI, LO, or dashes on the display | Target outside range, or meter too hot or cold | Check the range; let the meter acclimatise |
| Er or error code | Internal fault or operating outside limits | Fresh battery and room temperature, then the manual’s code list |
| Reading stays fixed after release | HOLD is the normal behaviour | Pull the trigger again for a new reading |
| Measures without holding the trigger | LOCK mode on | Turn LOCK off in the menu |
| Laser on but no reading | Weak battery, display fault | New battery; service if it persists |
| Readings vary with distance | Spot size grows with distance and includes background | Move closer, using the distance-to-spot ratio |
The distance-to-spot ratio matters a lot. A 12:1 meter measures a circle about 1 cm wide at 12 cm away and about 10 cm wide at 1.2 m. If the target is smaller than the spot, the meter averages in whatever is behind it.
Checking accuracy afterwards
The simplest home reference is an ice bath. Fill a glass to the top with crushed ice, add cold water to fill the gaps, stir and wait a minute. Aim the thermometer straight down at the surface of the water from close range so the spot falls entirely on the water, not the glass. With emissivity at 0.95, water reads close to 0 degrees C (32 F). Water has a high emissivity, close to 0.95 or slightly above, so a reading within a degree or two is typical for consumer meters, within the accuracy printed in the manual.
A second check is a surface whose temperature you know with a contact thermometer, such as a matte painted radiator panel or a strip of black electrical tape on a pipe, measured with both instruments at the same time. If the IR and contact readings agree within the stated accuracy, the meter is fine. Units with a calibration certificate, like the ennoLogic eT650D, come with a documented reference point, but the ice check is still useful before critical work. Our guide to choosing an infrared thermometer explains how accuracy specifications are usually written.
When it is a fault, and safety limits
Send the thermometer back or replace it if, after a fresh battery and a full settings reset, it fails the ice check by more than its stated accuracy, shows an error code it cannot clear, has a cracked or permanently fogged lens, or the display has missing segments. Clean the lens only with a soft cotton swab and a little water or isopropyl alcohol; solvents and rough cloths damage the coating.
Two safety points apply to every laser thermometer. Never point the laser at eyes, people or animals, or at reflective surfaces that bounce it back; even low-power pointer lasers can harm vision. And remember an IR thermometer measures only the surface: it will not show the core of food, the inside of a pipe, or whether an electrical part is live. For electrical inspection of panels and connections, a thermal camera gives a fuller picture; see thermal camera versus infrared thermometer. Most consumer IR thermometers are not meant for body temperature; use a medical thermometer for people.
Reset emissivity, then do the ice check
Swap in a fresh battery after a one-minute power drain, set emissivity back to 0.95, pick your units and switch off lock and alarm modes. Then aim at a stirred glass of ice water from close range. A reading near 0 degrees C means the thermometer is ready to use; anything far off means the lens or the meter needs attention.
Frequently Asked Questions
Does an infrared thermometer have a reset button?
Most do not. Removing the battery reboots the electronics, and settings such as emissivity and units are reset manually through the MODE or SET buttons.
What emissivity should my laser thermometer be set to?
0.95 is the usual default and suits most painted, matte, organic and non-metallic surfaces. Shiny metals have much lower emissivity and read low unless you use tape or paint on them.
How do I change a laser thermometer from Fahrenheit to Celsius?
Look for a C/F button, a slide switch inside the battery compartment, or a units option in the MODE menu. The instruction sheet shows which your model uses.
Why does my IR thermometer read wrong after a reset?
Check that emissivity really went back to 0.95, that the meter has reached room temperature, and that the target fills the measurement spot. A dirty lens also reads low.
Does the laser affect the temperature reading?
No. The laser only shows where you are aiming. The measurement comes from infrared energy collected by the lens over a spot that is usually larger than the laser dot.