A dew point sensor finds the temperature at which the water vapour in the air would start to condense. The reference method, a chilled mirror sensor, literally cools a small polished mirror until dew forms on it and measures the mirror’s temperature at that moment. Most everyday instruments do something simpler: a capacitive polymer sensor measures relative humidity, a thermometer measures air temperature, and the electronics calculate dew point from the two. Industrial sensors for very dry compressed air often use a metal oxide layer whose electrical properties change with tiny amounts of moisture.
What dew point actually is
Air at any temperature can hold only so much water vapour before it saturates. Cool a parcel of air without adding or removing water, and its relative humidity rises until it hits 100 percent; any further cooling forces water out as droplets. The temperature where that happens is the dew point. Below freezing, the vapour forms ice instead, and the equivalent figure is called the frost point.
Unlike relative humidity, dew point does not change when the air simply warms or cools. It tracks the actual amount of water in the air, which makes it a far better number for judging condensation on cold pipes, mould risk on walls, comfort in summer, or whether compressed air is dry enough for spraying paint. Our explainer on why dew point matters indoors covers the comfort and mould side.
A practical example makes the difference clear. Air at 22 C and 50 percent relative humidity has a dew point of roughly 11 C. Any surface colder than that, such as a cold water pipe, a window frame on a winter night or an uninsulated corner, will collect condensation. Heat the same air to 26 C and the relative humidity falls to around 40 percent, yet the dew point stays at 11 C, because the amount of water in the air has not changed. A dew point sensor therefore tells you directly which surfaces are at risk, whatever the room temperature does through the day.
The main sensor types compared
| Sensor | How it finds dew point | Where it is used | Strength and weakness |
|---|---|---|---|
| Chilled mirror | A thermoelectric cooler chills a mirror; a light and detector spot the first film of dew; a precise thermometer reads the mirror temperature | Calibration labs, meteorology references, critical processes | Measures dew point directly and very accurately; needs a clean mirror, slower and more expensive |
| Capacitive polymer RH plus temperature | A thin polymer absorbs water vapour, changing a capacitor’s value; combined with air temperature, dew point is calculated | Hygrometers, HVAC controls, weather stations, data loggers | Cheap, fast and small; accuracy falls at very high or very low humidity and with drift |
| Metal oxide (often aluminium oxide) | A porous oxide layer between two electrodes; adsorbed moisture changes its capacitance or impedance | Compressed air, gas lines, desiccant dryers, very low dew points | Handles very dry gas well; needs periodic recalibration and can drift |
| Saturated salt (lithium chloride) | A heated salt-coated element finds the temperature where the salt holds a steady moisture balance | Older HVAC and industrial installs | Simple and robust; slower and largely replaced by electronic types |
| Psychrometer (wet and dry bulb) | Evaporation cools a wet thermometer; the gap to the dry thermometer gives humidity and dew point via tables | Field checks, teaching, HVAC | No electronics needed; needs airflow and a clean, wet wick |
Inside the calculated reading
When a home or trade hygrometer shows dew point, it is nearly always calculated. The device reads relative humidity and air temperature, then applies an approximation such as the Magnus formula to work out the temperature at which that same amount of water vapour would saturate. The method is sound, but it carries the errors of both sensors forward.
That matters more than people expect. A small error in humidity becomes a noticeable dew point error, and near saturation a few percent RH can move the result by around a degree. At low humidity the opposite effect appears: the polymer sensor’s percentage accuracy limits how precisely very dry air can be described, which is why compressed air work uses oxide or mirror sensors instead. If you want to see the arithmetic, our step-by-step page on calculating dew point from temperature and humidity works through it.
Placement errors also pass straight through. If the temperature sensor sits in a warm housing or near a hand, the calculated dew point will be off even though the humidity sensor is fine.
How to check a dew point reading
- Let it settle. Give the instrument 15 to 30 minutes in the space before you trust it, longer if it has come from a very different temperature.
- Run a salt test on RH-based units. Seal the sensor in a bag or jar with a damp paste of table salt for several hours. Saturated sodium chloride holds the air near 75 percent RH at room temperature, so the reading should settle close to that. The full method is in how to calibrate a hygrometer.
- Cross-check with a second instrument. A digital unit that shows dew point alongside humidity and temperature, such as the ERAY digital psychrometer, gives a useful comparison, though it also relies on an electronic humidity sensor. A sling psychrometer with a wet wick is fully independent of electronic sensors. Compare in the same spot, with steady air.
- Use a cold glass check. Slowly add ice to water in a shiny metal cup while stirring with a thermometer. The temperature when the outside first fogs is a rough dew point. It will not be precise, but it catches gross errors.
- Send critical sensors away. For compressed air, cleanrooms and process gas, the only honest check is a calibration against a chilled mirror reference by a lab.
When the sensor is faulty or unsuited
Polymer humidity sensors drift with age, and faster after exposure to solvents, cleaning sprays, ammonia, or long periods of condensation. A dew point that reads high in a dry room, or sticks near the air temperature after a damp spell, suggests the humidity element is saturated or contaminated. Let it dry in warm, dry air; if it still fails a salt test, replace or recalibrate it.
Industrial oxide sensors that suddenly read wetter can be reporting a real problem, such as a failing dryer, saturated desiccant or a drain trap that is not working. Do not assume the sensor is wrong until you have checked the dryer. Mirror sensors misread when oils or salts coat the mirror; most have a cleaning or balancing routine for exactly this.
Finally, match the sensor to the job. A room hygrometer is fine for comfort and mould checks, but it is the wrong tool for proving that compressed air meets a very low dew point specification.
Do one salt test today
If your dew point comes from an ordinary hygrometer, run a salt test tonight: a damp salt paste in a sealed container, the sensor inside, and a reading after several hours. If it settles near 75 percent RH, your calculated dew point is trustworthy enough for condensation and mould decisions; if it does not, correct or replace the sensor before relying on it.
Frequently Asked Questions
Is dew point measured or calculated on most hygrometers?
Calculated. The instrument measures relative humidity and temperature, then works out dew point from them. Only chilled mirror instruments, and some industrial sensors, measure it directly.
What is the difference between dew point and frost point?
Dew point is where water vapour condenses as liquid; frost point is where it deposits as ice below freezing. For the same air, the frost point is slightly higher than the dew point would be, so sensors in very dry gas often state which one they report.
Why does my dew point reading change when I breathe near it?
Breath is warm and very humid, so it briefly raises both temperature and moisture around the sensor. Step back and let the reading settle for a few minutes.
What dew point is comfortable indoors?
Many people find dew points roughly in the 10 to 15 C range comfortable. Above about 18 C the air usually starts to feel muggy, and very low dew points feel dry on skin and eyes.
Can a dew point sensor be damaged by condensation?
Brief condensation usually does no lasting harm to a polymer sensor, but long periods soaked, or condensation carrying dirt and chemicals, can cause drift. Protective filters and dry storage help.